Sheared Epidural Catheter During an Elective Procedure
Authors:
- Karen NoblettKNKaren Noblett
- Amanda McKinneyAMAmanda McKinney
- Richard KimRKRichard Kim
- 1 February 2007
- journal article
- case report
- Published by Wolters Kluwer Health in Obstetrics & Gynecology
- Vol. 109 (2) , 566-568
Abstract
Epidural catheter fracture and retention is uncommon. Imaging techniques identifying catheter location are limited, and clear management guidelines are lacking. Radiculopathy and lumbar stenosis are recognized complications. During epidural placement in an elderly woman undergoing gynecologic surgery, a portion of the catheter was shorn off in the epidural space. Imaging, including plain X-ray and computed tomography scan had limited ability to identify the catheter. The initial neurosurgical consultation recommended leaving the catheter in place. However, the patient became symptomatic with radicular pain; 16.5 cm of the catheter were subsequently surgically removed. Retained epidural catheter fragments can be associated with long-term sequelae. Immediate removal is facilitated by a blood-clot tract and may present the best time for removal. Current imaging techniques have limited use in these situations.Keywords
This publication has 3 references indexed in Scilit:
- A Retained Racz?? Catheter Fragment After Epidural Neurolysis: Implications During Magnetic Resonance ImagingAnesthesia & Analgesia, 2003
- Radicular Pain Due to a Retained Fragment of Epidural CatheterAnesthesiology, 1997
- Lumbar Stenosis Complicating Retained Epidural Catheter TipAnesthesiology, 1995