Sensitivity to Pulse Rate and Amplitude Modulation in an Animal Model of the Auditory Brainstem Implant (ABI)
- 8 May 2023
- journal article
- research article
- Published by Springer Nature in Journal of the Association for Research in Otolaryngology
- Vol. 24 (3) , 365-384
Abstract
The auditory brainstem implant (ABI) is an auditory neuroprosthesis that provides hearing by electrically stimulating the cochlear nucleus (CN) of the brainstem. Our previous study (McInturff et al., 2022) showed that single-pulse stimulation of the dorsal (D)CN subdivision with low levels of current evokes responses that have early latencies, different than the late response patterns observed from stimulation of the ventral (V)CN. How these differing responses encode more complex stimuli, such as pulse trains and amplitude modulated (AM) pulses, has not been explored. Here, we compare responses to pulse train stimulation of the DCN and VCN, and show that VCN responses, measured in the inferior colliculus (IC), have less adaption, higher synchrony, and higher cross-correlation. However, with high-level DCN stimulation, responses become like those to VCN stimulation, supporting our earlier hypothesis that current spreads from electrodes on the DCN to excite neurons located in the VCN. To AM pulses, stimulation of the VCN elicits responses with larger vector strengths and gain values especially in the high-CF portion of the IC. Additional analysis using neural measures of modulation thresholds indicate that these measures are lowest for VCN. Human ABI users with low modulation thresholds, who score best on comprehension tests, may thus have electrode arrays that stimulate the VCN. Overall, the results show that the VCN has superior response characteristics and suggest that it should be the preferred target for ABI electrode arrays in humans.Keywords
Funding Information
- National Institute on Deafness and Other Communication Disorders (01089)
- Fondation Bertarelli
- Singeria Grant from the Swiss National Science Foundation (530733)
This publication has 56 references indexed in Scilit:
- Unanesthetized Auditory Cortex Exhibits Multiple Codes for Gaps in Cochlear Implant Pulse TrainsJournal of the Association for Research in Otolaryngology, 2011
- Anin vivoinvestigation of first spike latencies in the inferior colliculus in response to multichannel penetrating auditory brainstem implant stimulationJournal of Neural Engineering, 2010
- Selective Electrical Stimulation of the Auditory Nerve Activates a Pathway Specialized for High Temporal AcuityJournal of Neuroscience, 2010
- Inferior Colliculus Responses to Multichannel Microstimulation of the Ventral Cochlear Nucleus: Implications for Auditory Brain Stem ImplantsJournal of Neurophysiology, 2008
- Changes Across Time in Spike Rate and Spike Amplitude of Auditory Nerve Fibers Stimulated by Electric Pulse TrainsJournal of the Association for Research in Otolaryngology, 2007
- Effects of Stimulation Rate, Mode and Level on Modulation Detection by Cochlear Implant UsersJournal of the Association for Research in Otolaryngology, 2005
- What's a cerebellar circuit doing in the auditory system?Trends in Neurosciences, 2004
- Responses of dorsal cochlear nucleus single units to electrical pulse train stimulation of the auditory nerve with a cochlear implant electrodeThe Journal of the Acoustical Society of America, 1995
- Influence of neonatal cochlear removal on the development of mouse cochlear nucleus: I. Number, size, and density of its neuronsJournal of Comparative Neurology, 1982
- Cochlear nuclear complex of miceJournal of Anatomy, 1982