Dry Electrode Materials for Electrocardiographic Monitoring
Authors:
- 1 December 2018
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 645-646
Abstract
In this paper, a preliminary study about the most suitable dry electrode materials is done in the context of electrocardiogram (ECG) measurement systems used at home. Therefore, the electrical equivalent circuit between two electrodes through the body is developed as well as the ECG quality evaluation factors. Finally, an initial series of impedance and ECG quality measurements is presented which led to the conclusion that dry stainless steel electrodes seem to provide the best ECG signal quality.Keywords
This publication has 3 references indexed in Scilit:
- Normalization: A Preprocessing StageIARJSET, 2015
- Accurate Simulations of Electric Double Layer Capacitance of UltramicroelectrodesThe Journal of Physical Chemistry C, 2011
- CPE analysis by local electrochemical impedance spectroscopyElectrochimica Acta, 2006