Abstract
The present work presents the results obtained using an original open-source low-cost sensor (LCS) system developed to measure tropospheric O3 in a remote high altitude alpine site. We conducted our study at the Col Margherita Observatory (2543 m a.s.l.), a World Meteorological Organization Global Atmosphere Watch Regional Station, located in the Italian Eastern Alps. The sensing system mounts three equivalent commercial low-cost sensors that have been calibrated using a laboratory standard (Thermo 49iPS), referenced to the Standard Reference Photometer #15 calibration scale by the WMO, before field deployment. Intra and inter-comparison between sensors and reference (Thermo 49c) have been conducted for seven months from May to December 2018. The sensors required an individual calibration, both in laboratory and in the field. The sensor's dependence on the environmental meteorological variables has been considered and discussed. The evaluation of the analytical performances of this sensing system provides an LOD < 5 ppb, LOQ < 17 ppb, LDR up to 250 ppb, intra-Pearson correlation coefficient (PCC) up to 0.96, inter-PCC > 0.8, bias > 3.5 ppb and ±8.5 at 95 % of confidence. Thanks to the first implementation of an LCS System in an alpine site, it has been demonstrated how it is possible to obtain valuable data from a low-cost instrument in a remote environment. This opens new perspectives for the adoption of a low-cost sensor network in atmospheric sciences.