Diagnosis On-Demand: Field Evaluation of Microfluidic Paper Device for the Detection of Asymptomatic Malaria
- 12 May 2025
- journal article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 97 (22) , 11787-11797
Abstract
We report a proof-of-concept surveillance strategy that combines simplicity with modest cost to achieve high analytical performance for the detection of asymptomatic malaria infection. We developed a microfluidic paper-based analytical device (μPAD) that automates immunoassay using 30 μL of whole blood. A novel dendrimer-based signal transduction method was introduced to amplify the mass spectrometry (MS) signal. Our ability to detect an immunoassay signal with MS was achieved using cleavable ionic probes that were attached to monoclonal antibodies via dendrimer bioconjugation. The limit of detection of this MS-based immunoassay was determined to be 4.5 pM in serum for Plasmodium falciparum histidine-rich protein 2. The high stability of the ionic probes allowed storage of the paper device at room temperature, facilitating field sampling of whole blood from 266 asymptomatic volunteers in Ghana. The performance of the μPAD platform was compared with polymerase chain reaction (PCR), light microscopy, and rapid diagnostic test (RDT). Using PCR as the reference, we observed 96.5% sensitivity for our μPAD platform, as opposed to 43% for RDT and 16.9% for microscopy. Cohen's Kappa statistical analysis confirmed almost perfect agreement between μPAD and PCR (%agree = 81% in Kappa's categorical terms). False-negative rate in μPAD was calculated to be 3.5%. These results highlight the need for a more sensitive tool for asymptomatic malaria screening. Our method is highly sensitive yet deliverable, making sustainable implementation possible in underserved communities.Keywords
Funding Information
- National Institute of Allergy and Infectious Diseases (R01-AI 143809)
This publication has 26 references indexed in Scilit:
- Dendrimers as versatile platforms for advanced biosensor development: A review on chemistry, synthesis, and performance enhancementsJournal of Environmental Chemical Engineering, 2025
- Malaria Diagnosis Using Paper-Based Immunoassay for Clinical Blood Sampling and Analysis by a Miniature Mass SpectrometerAnalytical Chemistry, 2022
- Automated Immunoassay Performed on a 3D Microfluidic Paper-Based Device for Malaria Detection by Ambient Mass SpectrometryAnalytical Chemistry, 2022
- Empowering Clinical Diagnostics with Mass SpectrometryACS Omega, 2020
- An Update on Malaria Rapid Diagnostic TestsCurrent Infectious Disease Reports, 2018
- Mass Spectrometry for Paper-Based Immunoassays: Toward On-Demand DiagnosisJournal of the American Chemical Society, 2016
- From the Bench to the Field in Low‐Cost Diagnostics: Two Case StudiesAngewandte Chemie International Edition, 2015
- Immunohistochemistry and mass spectrometry for highly multiplexed cellular molecular imagingLaboratory Investigation, 2015
- A comparative laboratory diagnosis of malaria: microscopy versus rapid diagnostic test kitsAsian Pacific Journal of Tropical Biomedicine, 2012
- EVALUATION OF THE POLYMERASE CHAIN REACTION ANALYSIS FOR DIAGNOSIS OF FALCIPARUM MALARIA IN DELHI, INDIAIndian Journal of Medical Microbiology, 2005