The Effect of Implant Length and Diameter on Stress Distribution of Tooth-Implant and Implant Supported Fixed Prostheses: An In Vitro Finite Element Analysis Study
- 1 February 2023
- journal article
- research article
- Published by American Academy of Implant Dentistry in Journal of Oral Implantology
- Vol. 49 (1) , 46-54
Abstract
The aim of this study is to evaluate the effectiveness of the implant diameter and length on force dissemination of tooth-implant and implant retained fixed restorations. A finite analysis model was used via a 3D simulation of a unilateral mandibular Kennedy Class I arch. Through thresholding the resultant assembly, a region of interest was selected from the computed tomography (CT) scan. Details of the diameter (D) and length (L) of implant were introduced. Ds used were 3.7, 4.7, and 5.7, while Ls used were 10, 11.5, and 13. The constant was the use of rigid connectors in both designs (implant-implant and implant-tooth fixed partial dentures [FPDs]) and the mesial implant (D 3.7 and L 11.5). Stress in cancellous bone around mesial abutment, which is the second premolar in tooth-implant FPD and mesial implant in the implant-implant FPD, revealed that the stress was significantly lower in tooth-implant FPD when compared with implant-implant FPD (21.1 ± 0.00 vs 46.1 ± 0.00, P < .001). Stress distribution in the bone around any implant depends on several factors such as diameter, length, and tooth-implant vs implant-implant support. The implant diameter was more significant for improved stress distribution than implant length. A moderate increase in the length of the implant consequently reduced stress.Keywords
This publication has 36 references indexed in Scilit:
- The effect of rigid and non‐rigid connections between implants and teeth on biological and technical complications: a systematic review and a meta‐analysisClinical Oral Implants Research, 2016
- The Overall Survival, Complication-Free Survival, and Related Complications of Combined Tooth-Implant Fixed Partial Dentures: A Literature ReviewDentistry Journal, 2016
- Stress analysis in bone tissue around single implants with different diameters and veneering materials: A 3-D finite element studyMaterials Science and Engineering: C, 2013
- Effect of Diameter and Length on Stress Distribution of the Alveolar Crest around Immediate Loading ImplantsClinical Implant Dentistry and Related Research, 2009
- Required minimum primary stability and torque values for immediate loading of mini dental implants: an experimental study in nonviable bovine femoral boneOral Surgery Oral Medicine Oral Pathology Oral Radiology and Endodontology, 2008
- Bivariate Evaluation of Cylinder Implant Diameter and Length: A Three‐Dimensional Finite Element AnalysisJournal of Prosthodontics, 2008
- Survival and complication rates of combined tooth–implant‐supported fixed partial denturesClinical Oral Implants Research, 2006
- Influence of Implant Length and Bicortical Anchorage on Implant Stress DistributionClinical Implant Dentistry and Related Research, 2003
- Three-dimensional finite element analysis of stress-distribution around single tooth implants as a function of bony support, prosthesis type, and loading during functionThe Journal of Prosthetic Dentistry, 1996
- Finite Element Stress Analysis of Tooth‐to‐Implant Fixed Partial Denture DesignsJournal of Prosthodontics, 1993