Prediction of scratch resistance of cobalt chromium alloy bearing surface, articulating against ultra-high molecular weight polyethylene, due to third-body wear particles
Authors:
- 1 January 2004
- journal article
- research article
- Published by SAGE Publications in Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
- Vol. 218 (1) , 41-50
Abstract
The entrapment of abrasive particles within the articulation between a cobalt chromium alloy (CoCrMo) femoral component and an ultra-high molecular weight polyethylene (UHMWPE) cup of artificial hip joints or tibial inserts of artificial knee joints usually scratches the metallic bearing surface and consequently increases the surface roughness. This has been recognized as one of the main causes of excessive polyethylene wear, leading to osteolysis and loosening of the prosthetic components. The purpose of this study was to use the finite element method to investigate the resistance of the cobalt chromium alloy bearing surface to plastic deformation, as a first approximation to causing scratches, due to various entrapped debris such as bone, CoCrMo and ZrO 2 (contained in radiopaque polymethyl methacrylate cement). A simple axisymmetric micro contact mechanics model was developed, where a spherical third-body wear particle was indented between the two bearing surfaces, modelled as two solid cylinders of a given diameter, under the contact pressure determined from macro-models representing either hip or knee implants. The deformation of both the wear particle and the bearing surfaces was modelled and was treated as elastic-plastic. The indented peak-to-valley height on the CoCrMo bearing surface from the finite element model was found to be in good agreement with that reported in a previous study when the third-body wear particle was assumed to be rigid. Under the physiological contact pressure experienced in both hip and knee implants, ZrO 2 wear particles were found to be fully embedded within the UHMWPE bearing surface, and the maximum von Mises stresses within the CoCrMo bearing surface reached the corresponding yield strength. Consequently, the CoCrMo bearing surface was deformed plastically and the corresponding peak-to-valley height (surface roughness) was found to increase with both the hardness and the size of the wear particle. Even in the case of CoCrMo wear particles, with similar mechanical properties to those of the CoCrMo bearing surface, a significant plastic deformation of the bearing surface was also noted; this highlighted the importance of considering the deformation of the wear particles. These findings support the hypotheses made by clinical studies on the contribution of entrapped debris to increased surface roughness of CoCrMo femoral bearing surfaces.Keywords
This publication has 23 references indexed in Scilit:
- The performance of gamma- and EtO-sterilised UHWMPE acetabular cups tested under severe simulator conditions. Part 1: role of the third-body wear processBiomaterials, 2002
- Modeling of damage to articulating surfaces by third body particles in total joint replacementsJournal of Materials Science: Materials in Medicine, 2000
- Influence of a foreign body on the wear of metallic femoral heads and polyethylene acetabular cups of total hip prosthesesJournal of Materials Science, 2000
- A hip joint simulator study using new and physiologically scratched femoral heads with ultra-high molecular weight polyethylene acetabular cupsProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 2000
- Third-body wear of cobalt-chromium-molybdenum implant alloys initiated by bone and poly(methyl methacrylate) particlesJournal of Biomedical Materials Research, 2000
- The prediction of polyethylene wear rate and debris morphology produced by microscopic asperities on femoral headsJournal of Materials Science: Materials in Medicine, 2000
- Biological reactions to wear debris in total joint replacementProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 2000
- Surface damage to cobalt-chrome femoral head prosthesesThe Journal of Bone and Joint Surgery. British volume, 1994
- Ceramic bearing surfaces in total artificial joints: Resistance to third body wear damage from bone cement particlesJournal of Medical Engineering & Technology, 1991
- The Influence of Bone and Bone Cement Debris on Counterface Roughness in Sliding Wear Tests of Ultra-High Molecular Weight Polyethylene on Stainless SteelProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1990