Testing a Cherenkov neutron time-of-flight detector on OMEGA
- 1 October 2018
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 89 (10) , 10I122
Abstract
A Cherenkov neutron time-of-flight (nTOF) detector developed and constructed at Lawrence Livermore National Laboratory was tested at 13 m from the target in a collimated line of sight (LOS) and at 5.3 m from the target in the open space inside the OMEGA Target Bay. Neutrons interacting with the quartz rod generate gammas, which through Compton scattering produce relativistic electrons that give rise to Cherenkov light. A photomultiplier tube (PMT) transferred the Cherenkov light into an amplified electrical signal. The Cherenkov nTOF detector consists of an 8-mm-diam, 25-cm quartz hexagonal prism coupled with a Hamamatsu gated PMT R5916U-52. The tests were performed with DT direct-drive implosions with cryogenic and room-temperature targets, producing a wide range of neutron yields and ion temperatures. The results of the tests and comparison with other nTOF detectors on OMEGA are presented. In the collimated LOS at 13 m from the target, the Cherenkov nTOF detector demonstrated good precision measurement in both the yield and ion temperature for DT yields above 3 × 1013.Keywords
Funding Information
- U.S. Department of Energy (DE-NA0001944)
This publication has 7 references indexed in Scilit:
- Calibration of a neutron time-of-flight detector with a rapid instrument response function for measurements of bulk fluid motion on OMEGAReview of Scientific Instruments, 2018
- A fused silica Cherenkov radiator for high precision time-of-flight measurement of DT γ and neutron spectra (invited)Review of Scientific Instruments, 2018
- A new neutron time-of-flight detector for fuel-areal-density measurements on OMEGAReview of Scientific Instruments, 2014
- High-resolution spectroscopy used to measure inertial confinement fusion neutron spectra on Omega (invited)Review of Scientific Instruments, 2012
- The National Ignition Facility neutron time-of-flight system and its initial performance (invited)Review of Scientific Instruments, 2010
- Prototypes of National Ignition Facility neutron time-of-flight detectors tested on OMEGAReview of Scientific Instruments, 2004
- Interpretation of neutron time-of-flight signals from current-mode detectorsReview of Scientific Instruments, 1997