Monitor unit prediction model for wobbling proton therapy with ridge filters
Authors:
- 10 October 2021
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 48 (12) , 8107-8116
Abstract
Purpose We introduced an output factor (cGy/MU) prediction model for wobbling proton beams over the full range of proton energy, scatterer thickness, and the width of spread-out Bragg peak (SOBP). Materials and Methods From Dec. 2015 to Aug. 2020, 1990 wobbling proton fields were used to treat patients, where 1714 fields had a diameter smaller than 11 cm and 276 had a diameter between 11 cm and 16 cm, which were designated as small and middle wobbling radius cases, respectively. The output factor is defined as the ratio of proton absorbed dose at mid-depth of SOBP to monitor unit (MU). It depends dominantly on proton energy, scatterer thickness, and the width of SOBP. We established the prediction model using the polynomial fitting function and determined its coefficients for the small and middle wobbling radius cases. We evaluated the accuracy of our prediction model by calculating the difference of predicted and measured output factors. Results For the small wobbling radius cases, the mean value of the output factor difference was 0.22% with a standard deviation of 1.3%. For the middle wobbling radius cases, the mean value was 0.20% and with a standard deviation of 0.79%. The large deviation was especially observed for wobbling proton beams having small field size and small width of SOBP. Conclusions We made a prediction model of output factor for wobbling proton beams, thereby determining monitor unit (MU) of each beam. This included the dependency of the output factor on the proton energy between 70 MeV to 230 MeV, scatterer thickness, and the width of SOBP. For 93.6% of the small and 95.5% of the middle wobbling radius cases, the deviation between predicted and measured output factor was below 3%. The cases with deviations of predicted and measured output factor above 3% had small field size and small width of SOBP. The accuracy of our prediction model would be improved by adopting the field size effect and measuring more cases of small field size and small SOBP width in the future. This article is protected by copyright. All rights reservedKeywords
Funding Information
- Fonds National de la Recherche Luxembourg (NRF‐2017R1D1A1B04030619)
This publication has 15 references indexed in Scilit:
- Implementation of an improved dose-per-MU model for double-scattered proton beams to address interbeamline modulation width variabilityJournal of Applied Clinical Medical Physics, 2014
- Commissioning of output factors for uniform scanning proton beamsMedical Physics, 2011
- Prediction of Output Factor, Range, and Spread-Out Bragg Peak for Proton TherapyMedical Dosimetry, 2010
- Field size dependence of the output factor in passively scattered proton therapy: Influence of range, modulation, air gap, and machine settingsMedical Physics, 2009
- An overview of the comprehensive proton therapy machine quality assurance procedures implemented at The University of Texas M. D. Anderson Cancer Center Proton Therapy Center–HoustonMedical Physics, 2009
- A procedure for calculation of monitor units for passively scattered proton radiotherapy beamsMedical Physics, 2008
- Evaluation of beam wobbling methods for heavy‐ion radiotherapyMedical Physics, 2008
- Determination of output factors for small proton therapy fieldsMedical Physics, 2007
- A compact ridge filter for spread out Bragg peak production in pulsed proton clinical beamsMedical Physics, 2001
- Wobbler facility for biomedical experimentsMedical Physics, 1987