A novel amplitude binning strategy to handle irregular breathing during 4DMRI acquisition: improved imaging for radiotherapy purposes
Authors:
- Z. van KesterenZKZ. van Kesteren
- A. van der HorstAHA. van der Horst
- O. J. Gurney-ChampionOGO. J. Gurney-Champion
- I. BonesIBI. Bones
- D. TekelenburgDTD. Tekelenburg
- T. AlderliestenTAT. Alderliesten
- G. van TienhovenGTG. van Tienhoven
- R. KlaassenRKR. Klaassen
- H. W. M. van LaarhovenHLH. W. M. van Laarhoven
- A. BelABA. Bel
Open Access
- 14 May 2019
- journal article
- research article
- Published by Springer Nature in Radiation Oncology
- Vol. 14 (1) , 80
Abstract
For radiotherapy of abdominal cancer, four-dimensional magnetic resonance imaging (4DMRI) is desirable for tumor definition and the assessment of tumor and organ motion. However, irregular breathing gives rise to image artifacts. We developed a outlier rejection strategy resulting in a 4DMRI with reduced image artifacts in the presence of irregular breathing. We obtained 2D T2-weighted single-shot turbo spin echo images, with an interleaved 1D navigator acquisition to obtain the respiratory signal during free breathing imaging in 2 patients and 12 healthy volunteers. Prior to binning, upper and lower inclusion thresholds were chosen such that 95% of the acquired images were included, while minimizing the distance between the thresholds (inclusion range (IR)). We compared our strategy (Min95) with three commonly applied strategies: phase binning with all images included (Phase), amplitude binning with all images included (MaxIE), and amplitude binning with the thresholds set as the mean end-inhale and mean end-exhale diaphragm positions (MeanIE). We compared 4DMRI quality based on: Data included (DI); percentage of images remaining after outlier rejection. Reconstruction completeness (RC); percentage of bin-slice combinations containing at least one image after binning. Intra-bin variation (IBV); interquartile range of the diaphragm position within the bin-slice combination, averaged over three central slices and ten respiratory bins. IR. Image smoothness (S); quantified by fitting a parabola to the diaphragm profile in a sagittal plane of the reconstructed 4DMRI. Data included (DI); percentage of images remaining after outlier rejection. Reconstruction completeness (RC); percentage of bin-slice combinations containing at least one image after binning. Intra-bin variation (IBV); interquartile range of the diaphragm position within the bin-slice combination, averaged over three central slices and ten respiratory bins. IR. Image smoothness (S); quantified by fitting a parabola to the diaphragm profile in a sagittal plane of the reconstructed 4DMRI. A two-sided Wilcoxon’s signed-rank test was used to test for significance in differences between the Min95 strategy and the Phase, MaxIE, and MeanIE strategies. Based on the fourteen subjects, the Min95 binning strategy outperformed the other strategies with a mean RC of 95.5%, mean IBV of 1.6 mm, mean IR of 15.1 mm and a mean S of 0.90. The Phase strategy showed a poor mean IBV of 6.2 mm and the MaxIE strategy showed a poor mean RC of 85.6%, resulting in image artifacts (mean S of 0.76). The MeanIE strategy demonstrated a mean DI of 85.6%. Our Min95 reconstruction strategy resulted in a 4DMRI with less artifacts and more precise diaphragm position reconstruction compared to the other strategies. Volunteers: protocol W15_373#16.007; patients: protocol NL47713.018.14Keywords
This publication has 35 references indexed in Scilit:
- Interfractional variability of respiration-induced esophageal tumor motion quantified using fiducial markers and four-dimensional cone-beam computed tomographyRadiotherapy and Oncology, 2017
- Addition of MRI for CT-based pancreatic tumor delineation: a feasibility studyActa Oncologica, 2017
- Considerable interobserver variation in delineation of pancreatic cancer on 3DCT and 4DCT: a multi-institutional studyRadiation Oncology, 2017
- Revisiting the Potential of Alternating Repetition Time Balanced Steady-State Free Precession Imaging of the Abdomen at 3 TInvestigative Radiology, 2016
- High-Quality T2-Weighted 4-Dimensional Magnetic Resonance Imaging for Radiation Therapy ApplicationsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2015
- Optimizing 4-Dimensional Magnetic Resonance Imaging Data Sampling for Respiratory Motion Analysis of Pancreatic TumorsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2015
- Differences in respiratory-induced pancreatic tumor motion between 4D treatment planning CT and daily cone beam CT, measured using intratumoral fiducialsActa Oncologica, 2014
- Four‐dimensional magnetic resonance imaging (4D‐MRI) using image‐based respiratory surrogate: A feasibility studyMedical Physics, 2011
- Phase and amplitude binning for 4D-CT imagingPhysics in Medicine & Biology, 2007
- Interobserver Delineation Variation Using CT versus Combined CT + MRI in Intensity–Modulated Radiotherapy for Prostate CancerStrahlentherapie und Onkologie, 2005