A steady flow analysis on the stented and non-stented sidewall aneurysm models
Authors:
- S.C.M YuSYS.C.M Yu
- J.B ZhaoJZJ.B Zhao
- 1 April 1999
- journal article
- Published by IOP Publishing in Medical Engineering & Physics
- Vol. 21 (3) , 133-141
Abstract
As part of a general investigation on the effects of blood flow patterns in sidewall aneurysm, in vitro steady flow studies on rigid aneurysm models have been conducted using Particle Image Velocimetry over a range of Reynolds number from 200 to 1600. Above Reynolds number 700, one large recirculating vortex would be formed, occupying the entire aneurysmal pouch. The centre of the vortex is located at region near to the distal neck. A pair of counter rotating vortices would however be formed at Reynolds numbers below 700. For all the aneurysm models considered, the vortex strength, in general, is stronger at higher Reynolds numbers but lower at larger aneurysm size. Maximum strength of the vortex is about 15% of the bulk mean velocity in the upstream parent tube. Estimates of the wall shear stresses are derived from the near wall velocity measurements. Highest level of wall shear stresses always appears at the distal neck of the aneurysmal pouch. Stents and springs of different porosity have been used to dampen the flow movement inside the aneurysm so as to induce the possible formation of thrombosis. It is found that the flow movement inside the aneurysmal pouch can be suppressed to less than 5% of the bulk mean velocity by both devices. Furthermore, regions of high wall shear stresses at the distal neck could also be suppressed by almost 90%. The present results would be useful for further improvements in stent (or spring) technology.Keywords
This publication has 12 references indexed in Scilit:
- Modeling of Flow in a Straight Stented and Nonstented Side Wall Aneurysm ModelJournal of Biomechanical Engineering, 1997
- In vitro study of haemodynamics in a giant saccular aneurysm model: influence of flow dynamics in the parent vessel and effects of coil embolisationNeuroradiology, 1994
- Three-Dimensional Simulation of Blood Flow in an Abdominal Aortic Aneurysm—Steady and Unsteady Flow CasesJournal of Biomechanical Engineering, 1994
- Steady Flow in Abdominal Aortic Aneurysm ModelsJournal of Biomechanical Engineering, 1993
- Thrombosis of difficult intracranial aneurysms by the endovascular placement of platinum-Dacron microcoilsJournal of Neurosurgery, 1992
- Haemodynamic stress in lateral saccular aneurysmsActa Neurochirurgica, 1987
- Basic flow structure in saccular aneurysms: A flow visualization studyHeart and Vessels, 1987
- Use of detachable balloons for proximal artery occlusion in the treatment of unclippable cerebral aneurysmsJournal of Neurosurgery, 1987
- Clinicopathological study of cerebral aneurysmsJournal of Neurosurgery, 1978
- Acute Vascular Endothelial Changes Associated with Increased Blood Velocity GradientsCirculation Research, 1968