Comparison Between Different Extraction (LLE and SPE) and Determination (HPLC and Capillary‐LC) Techniques in the Analysis of Selected PAHs in Water Samples
- 1 November 2005
- journal article
- research article
- Published by Taylor & Francis in Journal of Liquid Chromatography & Related Technologies
- Vol. 28 (19) , 3045-3056
Abstract
In this work, the extraction of 9 out of 16 PAHs pollutants according to US Environmental Protection Agency (EPA) procedures, was studied through liquid‐liquid extraction (LLE) and solid‐phase extraction (SPE). The analysis of PAHs was made by high performance liquid chromatography (HPLC), using both a Supelcosil LC 18 (25 cm×4.6 mm, 5 µm) column operating in the conventional HPLC mode and a capillary column (20 cm×0.25 mm, 5 µm), packed in house with Spherisorb ODS‐2 particles and operating in the capillary liquid chromatography (c‐LC) mode. Of the extraction techniques used, LLE revealed itself to be efficient in the extraction of the higher‐molecular‐weight PAHs, while SPE was adequate for the extraction of all PAHs. HPLC revealed to be more sensitive than c‐LC in the detection of PAHs in the sample concentration. However, since in c‐LC the dilution of the compounds in the mobile phase is less, the mass sensitivity was significantly higher than that obtained with conventional HPLC (that is important when a limited sample amount is available). In the real water samples analyzed no PAH was found under the analytical conditions used.Keywords
This publication has 13 references indexed in Scilit:
- State of the Art in Miniaturized Separation TechniquesCritical Reviews in Analytical Chemistry, 2002
- Trace analysis by combined use of off-line solid-phase extraction, on-column sample focusing and U-shape flow cell in capillary liquid chromatographyJournal of Chromatography A, 1998
- Microcolumn liquid chromatography: instrumentation, detection and applicationsJournal of Chromatography A, 1997
- Instrumental Requirements for Nanoscale Liquid ChromatographyAnalytical Chemistry, 1996
- Determination of polycyclic aromatic hydrocarbons by liquid chromatographyJournal of Chromatography A, 1993
- Spectral atlas of polycyclic aromatic compounds. Vol. II: (including data on physico-chemical properties, occurrence and biological activity). Edited by W. Karcher. D. Reidel Publishing Co (Kluwer Academic Publishers Group), 1988. Pp. 864, ISBN 0 7923 0008 4. Price: £129·00Environmental Pollution, 1989
- Recent Advances in Microcolumn Liquid ChromatographyAnalytical Chemistry, 1988
- Recent advances in microcolumn liquid chromatographyAnalytical Chemistry, 1988
- Modern analytical methods for environmental polycyclic aromatic compoundsChemical Society Reviews, 1981
- Rapid analysis of ribonucleosides and bases at the picomole level using pellicular cation exchange resin in narrow bore columnsAnalytical Chemistry, 1969