Failure to Detect an Unusual Obstruction in a Reinforced Endotracheal Tube with Fiberoptic Examination
Authors:
- Matthias PaulMPMatthias Paul
- Michael DueckMDMichael Dueck
- Sandra KampeSKSandra Kampe
- Frank PetzkeFPFrank Petzke
- 1 September 2003
- journal article
- case report
- Published by Wolters Kluwer Health in Anesthesia & Analgesia
- Vol. 97 (3) , 909-910
Abstract
Obstruction of an endotracheal tube (ETT) is a potentially life-threatening event. We report an unusual obstruction of a reinforced ETT. The valve-like obstruction was caused by a partial detachment of the inner coating from the embedded spiral of the ETT. It led to an increase in inspiratory airway pressure, failure to detect end-expiratory CO(2), and generated a wheezing sound in forced expiration. Fiberoptic inspection, which is a recommended procedure for a suspected ETT-obstruction, failed to identify this detachment. Exchanging the defective ETT immediately resolved the critical clinical situation. The detachment was most likely caused by re-autoclavation of the ETT, which was a specified single-use product.Keywords
This publication has 8 references indexed in Scilit:
- An unusual case of tracheal tube obstructionAnaesthesia, 2001
- Problems Related to the Endotracheal Tube: An Analysis of 2000 Incident ReportsAnaesthesia and Intensive Care, 1993
- An unusual case of endotracheal tube cuff dysfunctionJournal of Clinical Anesthesia, 1993
- CorrespondenceAnaesthesia and Intensive Care, 1993
- Pharyngeal Obstruction of a Reinforced Orotracheal TubeAnesthesia & Analgesia, 1993
- Obstruction of armoured tracheal tubes: case report and discussionCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1988
- Endotracheal Tube Obstruction Secondary to Oral Preoperative MedicationAnesthesia & Analgesia, 1984