Simulation of toric intraocular lens results: Manual keratometry versus dual-zone automated keratometry from an integrated biometer
Authors:
- Robert OsherRORobert Osher
- David CookeDCDavid Cooke
- Kerry SolomonKSKerry Solomon
- Helga SandovalHSHelga Sandoval
- Rosario Salas-CervantesRSRosario Salas-Cervantes
- Richard PotvinRPRichard Potvin
- 1 December 2011
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 37 (12) , 2181-2187
Abstract
To evaluate simulated clinical outcomes in patients with toric intraocular lenses (IOLs) calculated on the basis of dual-zone automated keratometry from an integrated optical biometer, relative to manual keratometry. Private practice, Mesa, Arizona, USA. Comparative case series. Patient records at 4 clinical sites were reviewed to identify patients who had manual keratometry and biometry with the Lenstar LS 900 recorded before toric IOL implantation and refractive follow-up data after implantation. Preoperative and operative data were extracted from patient charts. Simulated refractive outcomes were calculated based on mathematically removing the actual IOL implanted and then mathematically inserting the IOLs as determined by manual or automated keratometry from the biometry device. Data for 128 patients were available for analysis. The actual residual astigmatism was comparable between manual keratometry and automated keratometry from the biometry system. Although simulated residual refractive astigmatism was similar between the 2 devices on average, there was variability in results by patient. Simulated residual refractive astigmatism was lower for the biometer when the standard deviation of the angle of astigmatism was low. Site-to-site variability was lower with the biometer than with manual keratometry. Simulated outcomes suggest that overall results for a group of patients whose toric IOL surgery planning is performed with the dual-zone automated keratometry data from the biometer will be equivalent to those when manual keratometry is used. The reduced site-to-site variability with the biometer suggests an operational advantage.Keywords
This publication has 4 references indexed in Scilit:
- Comparison of the ultrasonographic method with 2 partial coherence interferometry methods for intraocular lens power calculationOptometry - Journal of the American Optometric Association, 2011
- Comparison of portable automated keratometry and manual keratometry for IOL calculationJournal of Cataract & Refractive Surgery, 1997
- Toric intraocular lenses: Correcting astigmatism while controlling axis shiftJournal of Cataract & Refractive Surgery, 1994
- Clinical comparison of automated and manual keratometry in pre-operative ocular biometryEye, 1992