Electrically Evoked Auditory Brain Stem Responses for Lateral and Medial Placement of the Clarion HiFocus Electrode
Authors:
- Jill B. FirsztJFJill B. Firszt
- Phillip A. WackymPWPhillip A. Wackym
- Wolfgang GagglWGWolfgang Gaggl
- Linda S. BurgLBLinda S. Burg
- Ruth M. ReederRRRuth M. Reeder
- 1 April 2003
- journal article
- Published by Wolters Kluwer Health in Ear & Hearing
- Vol. 24 (2) , 184-190
Abstract
The purpose of this study was to compare the electrically evoked auditory brain stem response (EABR) for lateral and medial placement of the Clarion HiFocus cochlear implant electrode array via the electrode positioning system (EPS). Twenty-five adult and pediatric cochlear implant recipients participated in the study. Intraoperatively recorded EABRs were evoked by stimuli via three intracochlear electrodes representing apical, medial, and basal locations, and responses were elicited before and after positioner insertion. Evoked potential measures of wave V amplitude and threshold were examined for statistical significance using ANOVA for repeated measures and Chi-Square methods. For a given supra-threshold stimulus level, the increase in EABR wave V amplitude was significantly larger after EPS placement compared to before EPS placement for electrodes 1 (apical) and 13 (basal). Likewise, when the stimulus was decreased to obtain a minimal amplitude, the wave V threshold was significantly lower after EPS placement for electrodes 7 (medial) and 13. The number of measurements that showed decreased wave V threshold after EPS insertion was significantly dependent on intracochlear electrode location. Placement of the Clarion Electrode Positioning System following HiFocus electrode insertion resulted in a reduction in the electrical current required to activate the auditory system. The effect of the EPS was greatest for the basal location, demonstrated by lower wave V thresholds and a larger percentage increase in wave V amplitude. The EABR reflected electrophysiologic changes relative to lateral-to-medial changes in intracochlear electrode position due to the EPS.Keywords
This publication has 13 references indexed in Scilit:
- Neurophysiology of Cochlear Implant Users I: Effects of Stimulus Current Level and Electrode Site on the Electrical ABR, MLR, and N1-P2 ResponseEar & Hearing, 2002
- Initial Evaluation of the Clarion CII Cochlear Implant: Speech Perception and Neural Response ImagingEar & Hearing, 2002
- Effects of the Clarion Electrode Positioning System on Auditory Thresholds and Comfortable Loudness Levels in Pediatric Patients With Cochlear ImplantsJAMA Otolaryngology–Head & Neck Surgery, 2001
- The Importance of Human Cochlear Anatomy for the Results of Modiolus-Hugging Multichannel Cochlear ImplantsOtology & Neurotology, 2001
- Relationship between Eabr Thresholds and Levels Used to Program the Clarion® Speech ProcessorAnnals of Otology, Rhinology & Laryngology, 1999
- Electrically Evoked Potentials Recorded in Adult and Pediatric Clarion® Implant UsersAnnals of Otology, Rhinology & Laryngology, 1999
- The recognition of sentences in noise by normal-hearing listeners using simulations of cochlear-implant signal processors with 6–20 channelsThe Journal of the Acoustical Society of America, 1998
- Effects of noise and spectral resolution on vowel and consonant recognition: Acoustic and electric hearingThe Journal of the Acoustical Society of America, 1998
- Potential distributions and neural excitation patterns in a rotationally symmetric model of the electrically stimulated cochleaHearing Research, 1995
- Intraoperative and Postoperative Electrically Evoked Auditory Brain Stem Responses in Nucleus Cochlear Implant UsersEar & Hearing, 1994