How many wavelength channels do we need in Thomson scattering diagnostics?
Authors:
- 30 August 1999
- journal article
- letter
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 70 (9) , 3780-3781
Abstract
The necessary number of wavelength channels is investigated in incoherent Thomson scattering diagnostics. The measurement errors (σ) are evaluated for a realistic case and their lower limits are given in explicit forms. For a given number of wavelength channels, optimal channel allocation in the wavelength space is sought by minimizing the average relative errors. As a scale of the goodness of measurement, we introduce the ratio averaged over the electron temperature of interest. This quantity ranges from 0 to 1 and tends to 1 as the lowest achievable error is approached. If we conclude that is sufficient, then for a typical ruby laser system measuring 400–700 nm, seven channels is an adequate number for covering the electron temperature of 0.01–10 keV and five channels for 0.1–10 keV.