Comparative Analysis of Extraction Complications in Transvenous Versus Leadless Pacemakers
Open Access
- 6 May 2026
- journal article
- review article
- Published by Springer Nature in Cureus
- Vol. 18 (5) , e108392
Abstract
Pacemakers are cardiac implantable electronic devices (CIEDs) that deliver electrical impulses to regulate heart rhythm and manage conditions such as cardiac arrhythmias. While transvenous pacemakers (TP) have a generator with pacing leads, leadless pacemakers (LPs) are anchored within the right ventricular endocardium via tined or helical fixation. Device extractions may occur due to infection, dysfunction, or device upgrade. This study aims to analyze the incidence and severity of extraction complications between TPs and LPs. A comprehensive narrative review was performed through the PubMed database. The following key search terms and synonyms were incorporated into the search algorithm: “transvenous pacemaker,” “leadless pacemaker,” “pacemaker extraction,” “device infection,” and “bacterial complications”. This yielded 788 entries. After uploading the search results, articles that focused on pacemaker extractions were included, while animal studies, pediatric studies, and studies that were not peer-reviewed were excluded. After excluding 691 studies, 97 original articles on pacemaker extractions underwent full-text analysis. Both TP and LP extractions had high procedural success and low mortality rates. TPs were found to have procedural risks absent from leadless, namely, the development of arteriovenous fistula through subclavian access, tricuspid dysfunction, and atrial tearing. Examples of some differences between these studies include one LP study, which reported no change in mortality at a follow-up period of 18 months, and one TP study showed little change in mortality up to 24 months. Patients with TP and bacteremia have an increased risk of infection in the pocket and along the pacing leads. Infection of pacing leads has been shown to increase mortality and long-term complications. LPs are favored in many cases, for they minimize the incidence of these complications, have shorter extraction times, and have better recovery post-op in comparison to TPs. LPs are associated with fewer extraction complications due to their smaller size and minimally invasive procedural approach, eliminating the use of a generator pocket and transvenous leads. Due to the incorporation of case studies, this review’s limited sample size may decrease external validity. Future studies should collect large-scale data to further analyze the extraction complications between TP and LP patients.Keywords
This publication has 32 references indexed in Scilit:
- Management of pacemaker lead decubitus in an adult with repaired tetralogy of Fallot: A case reportInternational Journal of Surgery Case Reports, 2025
- Leadless Pacemakers: The “Leading Edge” of Quality of Life in Cardiac ElectrophysiologyCurrent Cardiology Reports, 2025
- Retrieval and replacement feasibility of 7-year-old implanted leadless pacemaker with tines fixationHeartRhythm Case Reports, 2023
- Successful avoidance of superior vena cava injury during transvenous lead extraction using a tandem femoral-superior approachHeart Rhythm, 2022
- Percutaneous Extraction for Misplacement of Pacemaker Leads Within the Coronary Artery and Left VentricleJACC: Case Reports, 2021
- Nanostim leadless pacemaker retrieval and simultaneous micra leadless pacemaker replacement: a single-center experienceJournal of Interventional Cardiac Electrophysiology, 2019
- Updated performance of the Micra transcatheter pacemaker in the real-world setting: A comparison to the investigational study and a transvenous historical controlHeart Rhythm, 2018
- Leadless pacemaker versus transvenous single-chamber pacemaker therapy: A propensity score-matched analysisHeart Rhythm, 2018
- A worldwide experience of the management of battery failures and chronic device retrieval of the Nanostim leadless pacemakerHeart Rhythm, 2017
- Less-Invasive Surgical Extraction of Problematic or Infected Permanent Transvenous Pacemaker SystemThe Annals of Thoracic Surgery, 2005