Monte Carlo and experimental determination of correction factors for gamma knife perfexion small field dosimetry measurements
- 5 September 2017
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 62 (18) , 7532-7555
Abstract
Detector-, field size- and machine-specific correction factors are required for precise dosimetry measurements in small and non-standard photon fields. In this work, Monte Carlo (MC) simulation techniques were used to calculate the k Q m s r , Q 0 f m s r , f r e f and k Q c l i n , Q m s r f c l i n , f m s r correction factors for a series of ionization chambers, a synthetic microDiamond and diode dosimeters, used for reference and/or output factor (OF) measurements in the Gamma Knife Perfexion photon fields. Calculations were performed for the solid water (SW) and ABS plastic phantoms, as well as for a water phantom of the same geometry. MC calculations for the k Q c l i n , Q m s r f c l i n , f m s r correction factors in SW were compared against corresponding experimental results for a subset of ionization chambers and diode detectors. Reference experimental OF data were obtained through the weighted average of corresponding measurements using TLDs, EBT-2 films and alanine pellets. k Q m s r , Q 0 f m s r , f r e f values close to unity (within 1%) were calculated for most of ionization chambers in water. Greater corrections of up to 6.0% were observed for chambers with relatively large air-cavity dimensions and steel central electrode. A phantom correction of 1.006 and 1.024 (breaking down to 1.014 from the ABS sphere and 1.010 from the accompanying ABS phantom adapter) were calculated for the SW and ABS phantoms, respectively, adding up to k Q m s r , Q 0 f m s r , f r e f corrections in water. Both measurements and MC calculations for the diode and microDiamond detectors resulted in lower than unit k Q c l i n , Q m s r f c l i n , f m s r ...Keywords
Funding Information
- State Scholarships Foundation of Greece
This publication has 67 references indexed in Scilit:
- Spectral distribution of particle fluence in small field detectors and its implication on small field dosimetryMedical Physics, 2017
- Role of the density, density effect and mean excitation energy in solid-state detectors for small photon fieldsPhysics in Medicine & Biology, 2017
- Monte Carlo calculated correction factors for the PTW microDiamond detector in the Gamma Knife‐Model CMedical Physics, 2016
- On the Monte Carlo simulation of small-field micro-diamond detectors for megavoltage photon dosimetryPhysics in Medicine & Biology, 2015
- Monte Carlo calculated and experimentally determined output correction factors for small field detectors in Leksell Gamma Knife Perfexion beamsPhysics in Medicine & Biology, 2015
- Detector to detector corrections: A comprehensive experimental study of detector specific correction factors for beam output measurements for small radiotherapy beamsMedical Physics, 2014
- Output correction factors for nine small field detectors in 6 MV radiation therapy photon beams: A PENELOPE Monte Carlo studyMedical Physics, 2014
- Small fields output factors measurements and correction factors determination for several detectors for a CyberKnife® and linear accelerators equipped with microMLC and circular conesMedical Physics, 2013
- A new formalism for reference dosimetry of small and nonstandard fieldsMedical Physics, 2008
- AAPM's TG‐51 protocol for clinical reference dosimetry of high‐energy photon and electron beamsMedical Physics, 1999