Comparison of vessel contrast measured with a scanning-beam digital x-ray system and an image intensifier/television system
Authors:
- Michael A. SpeidelMSMichael A. Speidel
- Brian P. WilfleyBWBrian P. Wilfley
- Joseph A. HeanueJHJoseph A. Heanue
- Timothy D. BettsTBTimothy D. Betts
- Michael S. Van LyselMLMichael S. Van Lysel
- 15 February 2001
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 28 (2) , 232-240
Abstract
Vessel contrast was measured in the fluoroscopic images produced by a scanning-beam digital x-ray (SBDX) system and an image intensifier/television (II/TV) based system. The SBDX system electronically scans a series of pencil x-ray beams across the patient, each of which is directed at a distant small-area detector array. The reduction in detected scatter achieved with this geometry was expected to provide an increase in image contrast. Vessel contrast was evaluated from images of a phantom containing iodinated tubes. The vessels were inserted into an acrylic stack to provide a patient-mimicking scattering medium. Vessel diameter ranged from 0.3 to 3.1 mm. Images were acquired at 100 kVp with the SBDX and II/TV systems and averaged to reduce x-ray noise. The II/TV system was operated in the 6-in. image intensifier mode with an anti-scatter grid. The increase in contrast in the SBDX images, expressed as a ratio of the measured SBDX and II/TV contrasts, ranged from 1.63 to 1.79 for individual vessels. This agreed well with a prediction of the contrast improvement ratio for this experiment, based on measurements of the scatter fraction, object-plane line spread functions, and consideration of the source spectrum and detector absorption properties. The predicted contrast improvement ratio for SBDX relative to II/TV images was 1.62 to 1.77.Keywords
This publication has 14 references indexed in Scilit:
- An accurate method for computer‐generating tungsten anode x‐ray spectra from 30 to 140 kVMedical Physics, 1997
- DQE(f) of four generations of computed radiography acquisition devicesMedical Physics, 1995
- Effects of undersampling on the proper interpretation of modulation transfer function, noise power spectra, and noise equivalent quanta of digital imaging systemsMedical Physics, 1995
- Comparison of scatter measurement techniques in digital fluoroscopyPhysics in Medicine & Biology, 1993
- A simple method for determining the modulation transfer function in digital radiographyIEEE Transactions on Medical Imaging, 1992
- Absolute diameter measurements of coronary arteries based on the first zero crossing of the Fourier spectrumMedical Physics, 1989
- Absorption and noise in cesium iodide x‐ray image intensifiersMedical Physics, 1983
- Toward a unified view of radiological imaging systems. Part II: Noisy imagesMedical Physics, 1977
- Absorption and noise in x-ray phosphorsJournal of Applied Physics, 1973
- Truncation errors in calculating the MTF of radiographic screen-film systems from the line spread functionPhysics in Medicine & Biology, 1972