A new “tool” for transcatheter atrial defect closure: The St. Jude SL2™ sheath
- 16 March 2012
- journal article
- pediatric congenital-heart-disease
- Published by Wiley in Catheterization and Cardiovascular Interventions
- Vol. 80 (2) , 177-181
Abstract
Background: Orienting the left atrial disc of an atrial septal defect (ASD) closure device parallel to the septum is important, particularly in cases of ASD with deficient retroaortic rim. Standard vascular sheaths can make this challenging. We report the use of an electrophysiologic sheath, the SL2™ St. Jude Vascular sheath (St. Jude Medical, St. Paul, MN) for closure of ASD.Methods: All patients undergoing ASD closure between July 2010 and February 2011 at The Children's Hospital at Montefiore in which the SL2 sheath was used were reviewed.Results: Nine patients underwent transcatheter ASD or PFO (1/9) closure with the SL2 sheath. Mean age was 22.7 years (range, 3–58); weight was 46.5 kg (range, 14.7–78.6). The Amplatzer ASD Occluder (AGA Medical Corp., Plymouth, MN) was used in all cases. The mean measurement of the ASDs via “stop‐flow” technique was 13.3 mm (range, 4–19). The mean device size was 13.9 mm (range, 5–20) or 105% the size of the defect. The retroaortic rim was deficient (<2 mm) in 3 of 9. ASD closure was achieved in 9 of 9. In all, the left atrial disk was brought toward the atrial septum parallel to the atrial septum, and repositioning or recapturing of the device was unnecessary in any case. Mean fluoroscopy time was 9.3 min (range, 2.2–18). No significant complications.Conclusions: The SL2 sheath was useful to position the ASD device parallel to the atrial septum, eliminating the need for redeployment during the procedure. This sheath was useful in those cases of deficient retroaortic rim in this study.Keywords
This publication has 9 references indexed in Scilit:
- The parallel wire technique for septal defect closureCatheterization and Cardiovascular Interventions, 2008
- Major Complications Associated with Transcatheter Atrial Septal Occluder Implantation: A Review of the Medical Literature and the Manufacturer and User Facility Device Experience (MAUDE) DatabaseCongenital Heart Disease, 2007
- Device closure of large atrial septal defects: technical considerationsJournal of Cardiovascular Medicine, 2007
- Atrial and Ventricular Septal Defects Can Safely Be Closed by Percutaneous InterventionJournal of Interventional Cardiology, 2005
- Efetividade clínica e segurança do tratamento percutâneo da comunicação interatrial tipo ostium secundum, com a prótese Amplatzer®Arquivos Brasileiros de Cardiologia, 2004
- Erosion of Amplatzer septal occluder device after closure of secundum atrial septal defects: Review of registry of complications and recommendations to minimize future riskCatheterization and Cardiovascular Interventions, 2004
- Novel technique to prevent prolapse of the Amplatzer septal occluder through large atrial septal defectCatheterization and Cardiovascular Interventions, 2003
- Comparison of transcatheter closure of secundum atrial septal defect using the Amplatzer septal occluder associated with deficient versus sufficient rimsThe American Journal of Cardiology, 2002
- StarFlex ASD Closure: Deployment, Techniques, EquipmentJournal of Interventional Cardiology, 2001