Electromagnetic Interference To Infusion Pumps From GSM Mobile Phones
Authors:
- P. BartoliniPBP. Bartolini
- M. FlorisMFM. Floris
- M. TriventiMTM. Triventi
- P. CianfanelliPCP. Cianfanelli
- G. ScavinoGSG. Scavino
- L. ProiettiLPL. Proietti
- V. BarbaroVBV. Barbaro
- 3 February 2007
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE) in 2018 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC)
- Vol. 4, 3515-3518
Abstract
Electromagnetic interference (EMI) to critical care medical devices has been reported by various groups. Previuos studies have demonstrated that infusion and syringe pumps are susceptible of false alarm buzzing and block, when exposed to various EMI sources. Whether these events may have clinical relevance is still debated. The risk of EMI depends on several factors such as phone emitted power, distance and carrier frequency. We investigated the EMI on infusion and siringe pumps from GSM phones at various distances and emitted powers. Malfunctions were observed in 4/7 infusion pumps and 1/4 syringe pumps exposed to mobiles at their maximum output, for distances as long as 30 cm. The maximum power not inducing any malfunction even at 0 cm distance was also determined. The selection of a proper maximum power class reduces significantly the risk of EMI. Such a function is already built in the GSM standard and thus represents one of the feasible solutions to the EMI problem in hospitals.Keywords
This publication has 6 references indexed in Scilit:
- Use of mobile phones in hospitalThe Lancet, 2003
- Use of mobile phones in hospital: time to lift the ban?The Lancet, 2003
- CHARACTERIZATION OF ELECTROMAGNETIC INTERFERENCE OF MEDICAL DEVICES IN THE HOSPITAL DUE TO CELL PHONESHealth Physics, 2002
- Safe Use of Mobile Phones in HospitalsHealth Physics, 2000
- Electromagnetic interference on medical equipment by low-power mobile telecommunication systemsIEEE Transactions on Electromagnetic Compatibility, 2000
- Interference to medical equipment from mobile phonesJournal of Medical Engineering & Technology, 1997