Evaluating the accuracy of geometrical distortion correction of magnetic resonance images for use in intracranial brain tumor radiotherapy
Authors:
- 1 November 2019
- journal article
- research article
- Published by VM Media Group sp. z o.o in Reports of Practical Oncology and Radiotherapy
- Vol. 24 (6) , 606-613
Abstract
Determine the 1) effectiveness of correction for gradient-non-linearity and susceptibility effects on both QUASAR GRID3D and CIRS phantoms; and 2) the magnitude and location of regions of residual distortion before and after correction. Using magnetic resonance imaging (MRI) as a primary dataset for radiotherapy planning requires correction for geometrical distortion and non-uniform intensity. Phantom Study: MRI, computed tomography (CT) and cone beam CT images of QUASAR GRID3D and CIRS head phantoms were acquired. Patient Study: Ten patients were MRI-scanned for stereotactic radiosurgery treatment. Correction algorithm: Two magnitude and one phase difference image were acquired to create a field map. A MATLAB program was used to calculate geometrical distortion in the frequency encoding direction, and 3D interpolation was applied to resize it to match 3D T1-weighted magnetization-prepared rapid gradient-echo (MPRAGE) images. MPRAGE images were warped according to the interpolated field map in the frequency encoding direction. The corrected and uncorrected MRI images were fused, deformable registered, and a difference distortion map generated. Maximum deviation improvements: GRID3D , 0.27 mm y-direction, 0.07 mm z-direction, 0.23 mm x-direction. CIRS, 0.34 mm, 0.1 mm and 0.09 mm at 20-, 40- and 60-mm diameters from the isocenter. Patient data show corrections from 0.2 to 1.2 mm, based on location. The most-distorted areas are around air cavities, e.g. sinuses. The phantom data show the validity of our fast distortion correction algorithm. Patient-specific data are acquired in <2 min and analyzed and available for planning in less than a minute.Keywords
Funding Information
- Siemens Medical
This publication has 22 references indexed in Scilit:
- Development and validation of a novel large field of view phantom and a software module for the quality assurance of geometric distortion in magnetic resonance imagingMagnetic Resonance Imaging, 2015
- A Simple and Efficient Methodology To Improve Geometric Accuracy in Gamma Knife Radiation Surgery: Implementation in Multiple Brain MetastasesInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2014
- Dosimetric consequences of prostate-based couch shifts on the precision of dose delivery during simultaneous IMRT irradiation of the prostate, seminal vesicles and pelvic lymph nodesPhysica Medica, 2014
- Evaluation of Image-guidance Strategies for Prostate CancerTCRT Express, 2013
- A two‐step scheme for distortion rectification of magnetic resonance imagesMedical Physics, 2009
- Quality assurance of 3D-CRT: Indications and difficulties in their applicationsCritical Reviews in Oncology/Hematology, 2009
- A novel method for comparing 3D target volume delineations in radiotherapyPhysics in Medicine & Biology, 2008
- Will IGRT live up to its promise?Acta Oncologica, 2008
- A complete distortion correction for MR images: I. Gradient warp correctionPhysics in Medicine & Biology, 2005
- An image fusion study of the geometric accuracy of magnetic resonance imaging with the Leksell stereotactic localization systemJournal of Applied Clinical Medical Physics, 2001