Urgent Exchange of a HeartMate II Left Ventricular Assist Device After Percutaneous Lead Fracture
Authors:
- Igor D. GregoricIGIgor D. Gregoric
- Rajko RadovancevicRRRajko Radovancevic
- William E. CohnWCWilliam E. Cohn
- Nichole McGuireNMNichole McGuire
- O H. FrazierOFO H. Frazier
- 1 September 2009
- journal article
- case report
- Published by Wolters Kluwer Health in ASAIO Journal
- Vol. 55 (5) , 523-524
Abstract
In four of our patients implanted with the HeartMate II left ventricular assist device (LVAD), fraying and fracture of the percutaneous lead's internal wires led to pump failure. The patients included two adolescent males (age, 15 and 17 years), one adult male (age, 78 years), and one adult female (age, 49 years). The damage was attributed to patient-related twisting or kinking of the lead. The patients had been supported by the LVAD for 6-18 months before the lead was damaged. In each case, audible and visual alarms were activated, and the pump temporarily continued to function well. The patients were rushed to the hospital, and the pumps were exchanged. All four patients recovered uneventfully and were discharged from the hospital 10-14 days later. They continued to do well on LVAD support for up to 639 days after the initial implant procedure. These events show the importance of educating LVAD patients to avoid excessive twisting or kinking of the device's driveline.Keywords
This publication has 6 references indexed in Scilit:
- Improved Survival and Decreasing Incidence of Adverse Events With the HeartMate II Left Ventricular Assist Device as Bridge-to-Transplant TherapyThe Annals of Thoracic Surgery, 2008
- Clinical Experience With Sternotomy Versus Subcostal Approach for Exchange of the HeartMate XVE to the HeartMate II Ventricular Assist DeviceThe Annals of Thoracic Surgery, 2008
- Exchange Techniques for Implantable Ventricular Assist DevicesASAIO Journal, 2008
- Small Pumps for Ventricular Assistance: Progress in Mechanical Circulatory SupportCardiology Clinics, 2007
- Use of a Continuous-Flow Device in Patients Awaiting Heart TransplantationNew England Journal of Medicine, 2007
- Incidence and clinical management of life-threatening left ventricular assist device failureThe Journal of Heart and Lung Transplantation, 2004