Large tube section is the key to successful coronary thrombus aspiration: Findings of a standardized bench test
Authors:
- 5 December 2005
- journal article
- research article
- Published by Wiley in Catheterization and Cardiovascular Interventions
- Vol. 67 (2) , 254-257
Abstract
Thrombus removal by aspiration is one of the adjunctive techniques used to avoid embolization during PCI for acute myocardial infarction. Numerous devices are now available, but little is known about the mechanical rationale used in comparing them. The aim of the present study was to determine parameters to obtain optimal thrombus aspiration (TA). Heparin‐ and antiplatelet‐free blood samples were aspirated into 3 mm diameter standardized glass tubes to create a 30 mm long thrombus. Thrombus formation took place at room temperature over a period of 6 or 12 hr. Various catheters were tested using a variable vacuum device: three with right‐angle distal tip (0.038″, 0.067″, and 0.070″) and one with a beveled distal (length of the beveled, 0.054″; inner diameter catheter, 0.040″). The single endpoint was complete thrombus aspiration. A total of 103 TAs were presented for the four catheters. For 6‐ or 12‐hr‐old thrombus for a given catheter, there was no significant difference in vacuum pressure required to succeed TA (P = 0.47). For 6‐ or 12‐hr‐old thrombus, the larger the contact area is, the lower the pressure needed to aspirate the thrombus. Moreover, a beveled distal tip length (0.054″) does not make it possible to succeed TA at a lower pressure. The main factor for successful TA for thrombi ≥ 6 hr is inner diameter and not immediate thrombus contact area.Keywords
This publication has 13 references indexed in Scilit:
- Improved performance of a new thrombus aspiration catheter: Outcomes from in vitro experiments and a case presentationCatheterization and Cardiovascular Interventions, 2004
- Usefulness of a 6 Fr Right Judkins Catheter for Mechanically Extracting a Massive Intracoronary Thrombus From an Ectasic Right Coronary Artery: A report on two different cases of thrombectomyJapanese Heart Journal, 2004
- Society of cardiac angiography and interventions: Suggested management of the no‐reflow phenomenon in the cardiac catheterization laboratoryCatheterization and Cardiovascular Interventions, 2003
- Export aspiration catheter thrombosuction before actual angioplasty in primary coronary intervention for acute myocardial infarctionCatheterization and Cardiovascular Interventions, 2002
- Plaque Gruel of Atheromatous Coronary Lesion May Contribute to the No-Reflow Phenomenon in Patients With Acute Coronary SyndromeCirculation, 2002
- Integrated Analysis of Myocardial Blush and ST-Segment Elevation Recovery After Successful Primary AngioplastyCirculation, 2002
- Incidence and clinical significance of distal embolization during primary angioplasty for acute myocardial infarctionEuropean Heart Journal, 2002
- Early Resolution of ST-Segment Elevation Correlates With Myocardial Salvage Assessed by Tc-99m Sestamibi Scintigraphy in Patients With Acute Myocardial Infarction After Mechanical or Thrombolytic Reperfusion TherapyCirculation, 2002
- Intracoronary Thrombectomy With the X-Sizer Catheter System Improves Epicardial Flow and Accelerates ST-Segment Resolution in Patients With Acute Coronary SyndromeCirculation, 2002
- A new thrombectomy catheter device (AngioJet) for the disruption of thrombi: An in vitro studyCatheterization and Cardiovascular Interventions, 1999