Measurement and 3D Visualization of the Human Internal Heat Field by Means of Microwave Radiometry
Authors:
- Igor Alexandrovich SidorovISIgor Alexandrovich Sidorov
- Alexsandr Grigorevich GudkovAGAlexsandr Grigorevich Gudkov
- Vitalij Yurievich LeushinVLVitalij Yurievich Leushin
- Eugenia Nikolaevna GorlachevaEGEugenia Nikolaevna Gorlacheva
- Eugenij Pavlovich NovichikhinENEugenij Pavlovich Novichikhin
- Svetlana Victorovna AgasievaSASvetlana Victorovna Agasieva
Open Access
Abstract
The possibility of non-invasive determination of the depth of the location and temperature of a cancer tumor in the human body by multi-frequency three-dimensional (3D) radiothermography is considered. The models describing the receiving of the human body’s own radiothermal field processes are presented. The analysis of the possibility of calculating the desired parameters based on the results of measuring antenna temperatures simultaneously in two different frequency ranges is performed. Methods of displaying on the monitor screen the three-dimensional temperature distribution of the subcutaneous layer of the human body, obtained as a result of data processing of a multi-frequency multichannel radiothermograph, are considered. The possibility of more accurate localization of hyperthermia focus caused by the presence of malignant tumors in the depth of the human body with multi-frequency volumetric radiothermography is shown. The results of the study of various methods of data interpolation for displaying the continuous intrinsic radiothermal field of the human body are presented. Examples of displaying the volumetric temperature distribution by the moving plane method based on digital models and the results of an experimental study of the thermal field of the human body and head are given.Keywords
This publication has 41 references indexed in Scilit:
- Prospects for Application of Radio-Frequency Identification Technology with Passive Tags in Invasive Biosensor SystemsBiomedical Engineering, 2015
- Stereoscopic three-dimensional visualization applied to multimodal brain images: clinical applications and a functional connectivity atlasFrontiers in Neuroscience, 2014
- Realistic Microwave Breast Models Through T1-Weighted 3-D MRI DataIEEE Transactions on Biomedical Engineering, 2014
- Toward wearable wireless thermometers for internal body temperature measurementsIEEE Communications Magazine, 2014
- Stable microwave radiometry system for long term monitoring of deep tissue temperature2013
- Five‐band microwave radiometer system for noninvasive brain temperature measurement in newborn babies: Phantom experiment and confidence intervalRadio Science, 2011
- NON-INVASIVE MICROWAVE RADIOMETRIC SYSTEM FOR INTRACRANIAL APPLICATIONS: A STUDY USING THE CONFORMAL L-NOTCH MICROSTRIP PATCH ANTENNAProgress In Electromagnetics Research, 2011
- Multifrequency Microwave Thermograph for Biomedical ApplicationsIEEE Transactions on Biomedical Engineering, 2004
- Multifrequency radiometric determination of temperature profiles in a lossy homogeneous phantom using a dual-mode antenna with integral water bolusIEEE Transactions on Microwave Theory and Techniques, 2002
- Monitoring of deep brain temperature in infants using multi-frequency microwave radiometry and thermal modellingPhysics in Medicine & Biology, 2001