Conductance catheter measurements of lumen area of stenotic coronary arteries: theory and experiment
Open Access
- 1 September 2011
- journal article
- Published by American Physiological Society in Journal of Applied Physiology
- Vol. 111 (3) , 758-765
Abstract
An injection of saline solution is required for the measurement of vessel lumen area using a conductance catheter. The injection of room temperature saline to displace blood in a vessel inevitably involves mass and heat transport and electric field conductance. The objective of the present study is to understand the accuracy of conductance method based on the phenomena associated with the saline injection into a stenotic blood vessel. Computational fluid dynamics were performed to simulate flow and its relation to transport and electric field in a stenotic artery for two different sized conductance catheters (0.9 and 0.35 mm diameter) over a range of occlusions [56–84% cross-sectional area (CSA) stenosis]. The results suggest that the performance of conductance catheter is dependent on catheter size and severity of stenosis more significantly for 0.9 mm than for 0.35 mm catheter. Specifically, the time of detection of 95% of injected saline solution at the detection electrodes was shown to range from 0.67 to 3.7 s and 0.82 to 0.94 s for 0.9 mm and 0.35 mm catheter, respectively. The results also suggest that the detection electrodes of conductance catheter should be placed outside of flow recirculation region distal to the stenosis to minimize the detection time. Finally, the simulations show that the accuracy in distal CSA measurements, however, is not significantly altered by whether the position of detection electrodes is inside or outside of recirculation zone (error was within 12% regardless of detection electrodes position). The results were experimentally validated for one lesion geometry and the simulation results are within 8% of actual measurements. The simulation of conductance catheter injection method may lead to further optimization of device and method for accurate sizing of diseased coronary arteries, which has clinical relevance to percutaneous intervention.Keywords
This publication has 27 references indexed in Scilit:
- Vortex formation and recirculation zones in left anterior descending artery stenoses: computational fluid dynamics analysisPhysics in Medicine & Biology, 2010
- A novel system for the reconstruction of a coronary artery lumen profile in real time: a preclinical validationAmerican Journal of Physiology-Heart and Circulatory Physiology, 2009
- Hemodynamic diagnostics of epicardial coronary stenoses: in-vitro experimental and computational studyBioMedical Engineering OnLine, 2008
- Calculation of arterial wall temperature in atherosclerotic arteries: effect of pulsatile flow, arterial geometry, and plaque structureBioMedical Engineering OnLine, 2007
- Novel method for measurement of medium size arterial lumen area with an impedance catheter: in vivo validationAmerican Journal of Physiology-Heart and Circulatory Physiology, 2005
- Three-dimensional finite element analysis of current density and temperature distributions during radio-frequency ablationIEEE Transactions on Biomedical Engineering, 1995
- Use of a conductance (volume) catheter and transient inferior vena caval occlusion for rapid determination of pressure‐volume relationships in manCatheterization and Cardiovascular Diagnosis, 1988
- Determination of left ventricular end-systolic pressure-volume relationships by the conductance (volume) catheter technique.Circulation, 1986
- In vivo demonstration of flow recirculation and turbulence downstream of graded stenoses in canine arteriesJournal of Biomechanics, 1985
- Continuous measurement of left ventricular volume in animals and humans by conductance catheter.Circulation, 1984