Validation of a new arterial pulse contour-based cardiac output device
Authors:
- Eric E. C. de WaalEWEric E. C. de Waal
- Cor J. KalkmanCKCor J. Kalkman
- Steffen RexSRSteffen Rex
- Wolfgang F. BuhreWBWolfgang F. Buhre
- 1 August 2007
- journal article
- clinical trial
- Published by Wolters Kluwer Health in Critical Care Medicine
- Vol. 35 (8) , 1904-1909
Abstract
To evaluate the accuracy and precision of an arterial pulse contour-based continuous cardiac output device (Vigileo). Vigileo cardiac output (VigileoCO) was compared with intermittent transpulmonary thermodilution cardiac output (TPCO) and an established arterial pulse contour-based cardiac output (PCCO). Prospective clinical study. University hospital. Twenty-two patients undergoing coronary artery bypass graft surgery. Defined volume load during surgery and in the postoperative period. We obtained 184 pairs of VigileoCO and TPCO, 140 pairs of VigileoCO and PCCO, and 140 pairs of PCCO and TPCO. Measurements were performed after induction of anesthesia (T1), after sternotomy (T2), immediately after (T3) and 20 mins after volume challenge with 10 mL·kg−1 hydroxyethyl starch 6% (T4), 15 mins after coronary pulmonary bypass (T5), after retransfusion of autologous blood (T6), after arrival at the intensive care unit (T7), and immediately after (T8) and 20 mins after (T9) a second volume load with 10 mL·kg−1 hydroxyethyl starch 6%. TPCO was used to calibrate PCCO. For pooled data, including uncalibrated PCCO data immediately after weaning from coronary pulmonary bypass (T5), the correlation coefficient of TPCO vs. VigileoCO, PCCO vs. VigileoCO, and TPCO vs. PCCO was 0.75, 0.60, and 0.75 respectively. Bland-Altman analysis showed a bias of 0.00, −0.01, and 0.02 L·min−1, the precision (=sd) was 0.87, 1.08, and 0.93 L·min−1, and the mean error was 33%, 40%, and 35%. When we compared calibrated PCCO values (T2–T4, T6, T7–9), the correlation coefficients of PCCO-VigileoCO and TPCO-PCCO were 0.72 and 0.85, bias was −0.16 and 0.19 L·min−1, and mean error was 33% and 27%, respectively. Best correlations and the least differences between TPCO and VigileoCO were observed in postbypass closed-chest conditions and in the intensive care unit. Our results showed that VigileoCO enables clinically acceptable assessment of cardiac output in postbypass closed-chest conditions and during stable conditions in the intensive care unit.Keywords
This publication has 18 references indexed in Scilit:
- Agreement between PiCCO pulse‐contour analysis, pulmonal artery thermodilution and transthoracic thermodilution during off‐pump coronary artery by‐pass surgeryActa Anaesthesiologica Scandinavica, 2006
- PulmonalarterienkatheterDer Anaesthesist, 2006
- Increased risk associated with pulmonary artery catheterization in the medical intensive care unitJournal of Critical Care, 2003
- Cardiac output monitoring: aortic transpulmonary thermodilution and pulse contour analysis agree with standard thermodilution methods in patients undergoing lung transplantationCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 2003
- Reliability of a new algorithm for continuous cardiac output determination by pulse-contour analysis during hemodynamic instabilityCritical Care Medicine, 2002
- Continuous cardiac output by femoral arterial thermodilution calibrated pulse contour analysis: Comparison with pulmonary arterial thermodilutionCritical Care Medicine, 1999
- Comparison of cardiac output assessed by pulse-contour analysis and thermodilution in patients undergoing minimally invasive direct coronary artery bypass graftingJournal of Cardiothoracic and Vascular Anesthesia, 1999
- A review of impedance cardiographyHeart & Lung, 1995
- Complications of Right Heart CatheterizationChest, 1985
- Catheterization of the Heart in Man with Use of a Flow-Directed Balloon-Tipped CatheterNew England Journal of Medicine, 1970