Evaluation of optical particulate matter sensors under realistic conditions of strong and mild urban pollution
Authors:
Open Access
- 30 November 2020
- journal article
- research article
- Published by Copernicus GmbH in Atmospheric Measurement Techniques
- Vol. 13 (12) , 6427-6443
Abstract
In this paper we evaluate characteristics of three optical particulate matter sensors/sizers (OPS): high-end spectrometer 11-D (Grimm, Germany), low-cost sensor OPC-N2 (Alphasense, United Kingdom) and in-house developed MAQS (Mobile Air Quality System), which is based on another low-cost sensor – PMS5003 (Plantower, China), under realistic conditions of strong and mild urban pollution. Results were compared against a reference gravimetric system, based on a Gemini (Dadolab, Italy), 2.3 m3 h−1 air sampler, with two channels (simultaneously measuring PM2.5 and PM10 concentrations). The measurements were performed in Sarajevo, the capital of Bosnia-Herzegovina, from December 2019 until May 2020. This interval is divided into period 1 – strong pollution – and period 2 – mild pollution. The city of Sarajevo is one of the most polluted cities in Europe in terms of particulate matter: the average concentration of PM2.5 during the period 1 was 83 µg m−3, with daily average values exceeding 500 µg m−3. During period 2, the average concentration of PM2.5 was 20 µg m−3. These conditions represent a good opportunity to test optical devices against the reference instrument in a wide range of ambient particulate matter (PM) concentrations. The effect of an in-house developed diffusion dryer for 11-D is discussed as well. In order to analyse the mass distribution of particles, a scanning mobility particle sizer (SMPS), which together with the 11-D spectrometer gives the full spectrum from nanoparticles of diameter 10 nm to coarse particles of diameter 35 µm, was used. All tested devices showed excellent correlation with the reference instrument in period 1, with R2 values between 0.90 and 0.99 for daily average PM concentrations. However, in period 2, where the range of concentrations was much narrower, R2 values decreased significantly, to values from 0.28 to 0.92. We have also included results of a 13.5-month long-term comparison of our MAQS sensor with a nearby beta attenuation monitor (BAM) 1020 (Met One Instruments, USA) operated by the United States Environmental Protection Agency (US EPA), which showed similar correlation and no observable change in performance over time.Keywords
All Related Versions
This publication has 48 references indexed in Scilit:
- Characterising low-cost sensors in highly portable platforms to quantify personal exposure in diverse environmentsAtmospheric Measurement Techniques, 2019
- Regression methods in the calibration of low-cost sensors for ambient particulate matter measurementsSN Applied Sciences, 2019
- Long-term field comparison of multiple low-cost particulate matter sensors in an outdoor urban environmentScientific Reports, 2019
- Evaluation of a Portable Aerosol Collector and Spectrometer to measure particle concentration by composition and sizeAerosol Science and Technology, 2019
- Evaluation of Low-Cost Sensors for Ambient PM2.5 MonitoringJournal of Sensors, 2018
- A thermodynamic description for the hygroscopic growth of atmospheric aerosol particlesAtmospheric Chemistry and Physics, 2018
- Precision and Accuracy of a Direct-Reading Miniaturized Monitor in PM2.5 Exposure AssessmentSensors, 2018
- Development of Low-Cost Air Quality Stations for Next Generation Monitoring Networks: Calibration and Validation of PM2.5 and PM10 SensorsSensors, 2018
- Evaluation of a low-cost optical particle counter (Alphasense OPC-N2) for ambient air monitoringAtmospheric Measurement Techniques, 2018
- Performance evaluation of the cost-effective and lightweight Alphasense optical particle counter for use onboard unmanned aerial vehiclesAerosol Science and Technology, 2017