Initial clinical evaluation of PET‐based ion beam therapy monitoring under consideration of organ motion
- 28 January 2016
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 43 (2) , 975-982
Abstract
Purpose: Intrafractional organ motion imposes considerable challenges to scanned ion beam therapy and demands for a thorough verification of the applied treatment. At the Heidelberg Ion‐Beam Therapy Center (HIT), the scanned ion beam delivery is verified by means of postirradiation positron‐emission‐tomography (PET) imaging. This work presents a first clinical evaluation of PET‐based treatment monitoring in ion beam therapy under consideration of target motion.Methods: Three patients with mobile liver lesions underwent scanned carbon ion irradiation at HIT and postirradiation PET/CT (x‐ray‐computed‐tomography) imaging with a commercial scanner. Respiratory motion was recorded during irradiation and subsequent image acquisition. This enabled a time‐resolved (4D) calculation of the expected irradiation‐induced activity pattern and, for one patient where an additional 4D CT was acquired at the PET/CT scanner after treatment, a motion‐compensated PET image reconstruction. For the other patients, PET data were reconstructed statically. To verify the treatment, calculated prediction and reconstructed measurement were compared with a focus on the ion beam range.Results: Results in the current three patients suggest that for motion amplitudes in the order of 2 mm there is no benefit from incorporating respiratory motion information into PET‐based treatment monitoring. For a target motion in the order of 10 mm, motion‐related effects become more severe and a time‐resolved modeling of the expected activity distribution can lead to an improved data interpretation if a sufficient number of true coincidences is detected. Benefits from motion‐compensated PET image reconstruction could not be shown conclusively at the current stage.Conclusions: The feasibility of clinical PET‐based treatment verification under consideration of organ motion has been shown for the first time. Improvements in noise‐robust 4D PET image reconstruction are deemed necessary to enhance the clinical potential.Keywords
Funding Information
- European Commission (EC) (241851)
- European Commission (241851)
- European Commission (241851)
This publication has 36 references indexed in Scilit:
- Analysis of metabolic washout of positron emitters produced during carbon ion head and neck radiotherapyMedical Physics, 2013
- Hypofractionated carbon ion therapy delivered with scanned ion beams for patients with hepatocellular carcinoma – feasibility and clinical responseRadiation Oncology, 2013
- Respiratory liver motion estimation and its effect on scanned proton beam therapyPhysics in Medicine & Biology, 2012
- Motion in radiotherapy: particle therapyPhysics in Medicine & Biology, 2011
- Monitoring proton radiation therapy with in-room PET imagingPhysics in Medicine & Biology, 2011
- Physical and clinical performance of the mCT time-of-flight PET/CT scannerPhysics in Medicine & Biology, 2011
- Phase i study evaluating the treatment of patients with hepatocellular carcinoma (HCC) with carbon ion radiotherapy: The PROMETHEUS-01 trialBMC Cancer, 2011
- Implementation and workflow for PET monitoring of therapeutic ion irradiation: a comparison of in-beam, in-room, and off-line techniquesPhysics in Medicine & Biology, 2011
- Variability of Four-Dimensional Computed Tomography Patient ModelsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008