Abstract
An electroanalytical method for determining dienestrol (DNL) in bovine urine samples is described. A glassy carbon electrode (GCE) modified with silver nanoparticles and functionalized multi-walled carbon nanotubes was used as working sensor. The modified GCE displays substantial analytical improvements including an amplified signal, fast electron transfer kinetics, and resistance to fouling. The irreversible oxidation signal of DNL is pH-dependent. Best reactivity is found at pH 3.0, where a typical anodic peak is recorded at 0.8 V (vs. Ag/AgCl). Square-wave voltammetry revealed a 8.4 nM detection limit (1.9 μg L−1), good repeatability and reproducibility (RSDs + and K+), urea and uric acid. The electrochemical sensor fulfills all requisites to be used as forensic device in surveillance of illegal livestock practices.
Funding Information
  • Coordination for the Improvement of Higher Education Personnel (88881.140821/2017-01)
  • Coordination for the Improvement of Higher Education Personnel (Finance code 001)
  • National Council for Scientific and Technological Development (420261/2018-4, 407891/2018-8)

This publication has 30 references indexed in Scilit: