Spectrophotometers intercomparison for spectrocolorimetric scale harmonization

Authors:

Abstract
An in situ comparison of spectrophotometer scales of wavelength, spectral bandwidth, and regular spectral transmittance was made in 1997/98 towards the harmonization of spectrocolorimetric scales for the measurements in transmittance mode. It used a set of three neutral density glass filter standards and a crystal with sharp absorption bands as wavelength and spectral bandwidth standard. Measurements by 22 laboratories with a great diversity of instrumentation constituted a good testing reference data to examine the aspects in scale harmonization. The intercomparison reflects a widespread misunderstanding about levels of accuracy achieved and a generalized lack of use of service standard coverage in checking uncertainty levels along the range of the concerned metrological scales. The spectral bandwidth scale is usually not taken in appropriated attention. It is demonstrated that relative small discrepancies in the spectral bandwidth setting lead to large uncertainties in the measurements that may overexceed computed uncertainty values derived from wavelength and spectral transmittance scale uncertainties.

This publication has 4 references indexed in Scilit: