Accuracy of smartwatches for measuring heart rate and oxygen saturation in cynomolgus macaques compared to clinical standards
Open Access
- 15 April 2025
- journal article
- Published by SAGE Publications in Laboratory Animals
- Vol. 59 (4) , 508-517
Abstract
Continuous monitoring of physiological parameters in non-human primates (NHPs) necessitates a precise, non-invasive, and convenient method. This study aimed to validate the use of smartwatches with integrated pulse oximetry and heart rate (HR) monitoring capabilities for use in NHPs. Currently, the clinical standard for non-invasive continuous monitoring of peripheral oxygen saturation (SpO 2 ) in NHPs has been the use of a transmittance pulse oximeter (TPO) affixed to a location of highly vascularized tissue. In a clinical setting, HR is monitored through electrocardiogram (ECG) or associated with SpO 2 measurement from a TPO probe utilizing photoplethysmography technology. Challenges in obtaining precise readings with TPOs stem from technological limitations and probe placement restrictions. To address these limitations, simultaneous HR and SpO 2 measurements were obtained from 15 cynomolgus macaques ( Macaca fascicularis ) using the Apple Watch 7 (AW 7), Apple Watch 9 (AW 9), and a clinical-grade TPO probe with integrated optical HR measurement technology (iM70, ELAN). Arterial blood gas (ABG) analysis was used as a reference method for SpO 2 . We found that a TPO device significantly underestimated SpO 2 compared to the AW 7 and AW 9 when referenced against ABG values. Smartwatch-derived HR and SpO 2 measurements demonstrated good agreement and minimal bias compared to the gold standard method. Overall, the AW 7 and AW 9 exhibited good agreement with clinical reference standards for HR and good agreement with the gold standard for SaO 2 in sedated cynomolgus macaques.Keywords
This publication has 29 references indexed in Scilit:
- Evaluating blood oxygen saturation measurements by popular fitness trackers in postoperative patients: A prospective clinical trialiScience, 2023
- Accuracy of the Apple Watch in measuring oxygen saturation: comparison with pulse oximetry and ABGIrish Journal of Medical Science (1971 -), 2023
- Accuracy of the Apple Watch Oxygen Saturation Measurement in Adults and Children with Congenital Heart DiseasePediatric Cardiology, 2022
- Comparison of SpO2 and heart rate values on Apple Watch and conventional commercial oximeters devices in patients with lung diseaseScientific Reports, 2021
- The accuracy of heartbeat detection using photoplethysmography technology in cardiac patientsJournal of Electrocardiology, 2021
- Investigating sources of inaccuracy in wearable optical heart rate sensorsnpj Digital Medicine, 2020
- Investigating the Effects of Temperature on PhotoplethysmographyIFAC-PapersOnLine, 2015
- Vascular complications and access crossover in 10,676 transradial percutaneous coronary proceduresAmerican Heart Journal, 2012
- Effect of Peripheral Vasoconstriction on Pulse OximetryJournal of Clinical Monitoring and Computing, 2006
- Peripheral vascular complications of coronary angioplasty by the femoral and brachial techniquesCatheterization and Cardiovascular Diagnosis, 1994