Comparative Biomechanical Evaluation of Different External Fixation Sidebars: Stainless-Steel Tubes Versus Carbon Fiber Rods
Authors:
- Emil H. SchemitschESEmil H. Schemitsch
- Richard M. HarringtonRHRichard M. Harrington
- Jens R. ChapmanJCJens R. Chapman
- Marc F. SwiontkowskiMSMarc F. Swiontkowski
- 1 October 1996
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Orthopaedic Trauma
- Vol. 10 (7) , 470-475
Abstract
Carbon fiber rods were developed to provide radiolucent sidebars for external fixation. In the present study, a single-plane, half-pin, double-bar external fixator construct with either stainless-steel tubes or carbon fiber rods was applied on the anteromedial surface of an osteotomized synthetic human tibia and evaluated for fixation rigidity. Testing was performed with the bone fragments in cortical contact and with a 5-mm midshaft gap between the fragments. The sidebars then were loaded to failure in bending. The results of this study show (a) that the carbon fiber rods were 15% stiffer than the stainless-steel tubes (p = 0.009) and (b) that the external fixator with carbon fiber rods achieved ≈85% of the fixation stiffness of the external fixator with stainless-steel tubes. The loss of stiffness of the external fixator with carbon fiber rods is most likely due to the clamps being less effective in connecting the carbon fiber rods rigidly to the Schanz screws.Keywords
This publication has 18 references indexed in Scilit:
- Treatment of closed tibial shaft fractures with unilateral external fixationInjury, 1995
- The Use of Combination Internal Fixation and Hybrid External Fixation in Severe Proximal Tibia FracturesJournal of Orthopaedic Trauma, 1995
- External fixation as a primary and definitive treatment of open limb fracturesInjury, 1995
- An Investigation of the Bending Stiffness of and the Plane Stresses Generated by a Flanged External Fixator PinJournal of Orthopaedic Trauma, 1995
- Bending stiffness of conical and standard external fixator pinsBiomaterials, 1993
- Guidelines for external fixation frame rigidity and stressesJournal of Orthopaedic Research, 1986
- Bending Stiffness of Unilateral and Bilateral Fixator FramesClinical Orthopaedics and Related Research, 1983
- Unilateral External Fixation for Severe Open Tibial FracturesClinical Orthopaedics and Related Research, 1983
- Development of a variable stiffness external fixation system for stabilization of segmental defects of the tibiaJournal of Orthopaedic Research, 1983