Abstract
We report the design and construction of a cavity enhanced absorption spectrometer using broadband Brewster’s angle prism retroreflectors and a spatially coherent 500 nm to > 1.75 µm supercontinuum excitation source. Using prisms made from fused silica an effective cavity reflectivity of > 99.99% at 1.064 µm was achieved. A proof of principle experiment was performed by recording the cavity enhanced absorption spectrum of the weak b-X (1←0) transition of molecular oxygen at 14529 cm-1 and the fifth overtone of the acetylene C-H stretch at 18430 cm-1. CCD frames were integrated for 150 sec and 30 sec, with 3 frames (each 100 cm-1 wide) and 1 frame (266 cm-1 wide) required to observe the O2 and C2H2 spectra, respectively. A rms noise equivalent absorption (αmin) of 7.21 × 10-8 cm-1 Hz-1/2 and 1.28 × 10-7 cm-1 Hz-1/2 with full width half maximum line widths of 0.18 cm-1 and 0.44 cm-1 were achieved for the molecular oxygen band and the acetylene overtone, respectively.