Simulation-based evaluation and optimization of a new CdZnTe gamma-camera architecture (HiSens)
Authors:
- Charlotte RobertCharlotte Robert
- Guillaume MontémontGMGuillaume Montémont
- Véronique RebuffelVRVéronique Rebuffel
- Irène BuvatIBIrène Buvat
- Lucie GuérinLGLucie Guérin
- Loïck VergerLVLoïck Verger
- 19 April 2010
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 55 (9) , 2709-2726
Abstract
A new gamma-camera architecture named HiSens is presented and evaluated. It consists of a parallel hole collimator, a pixelated CdZnTe (CZT) detector associated with specific electronics for 3D localization and dedicated reconstruction algorithms. To gain in efficiency, a high aperture collimator is used. The spatial resolution is preserved thanks to accurate 3D localization of the interactions inside the detector based on a fine sampling of the CZT detector and on the depth of interaction information. The performance of this architecture is characterized using Monte Carlo simulations in both planar and tomographic modes. Detective quantum efficiency (DQE) computations are then used to optimize the collimator aperture. In planar mode, the simulations show that the fine CZT detector pixelization increases the system sensitivity by 2 compared to a standard Anger camera without loss in spatial resolution. These results are then validated against experimental data. In SPECT, Monte Carlo simulations confirm the merits of the HiSens architecture observed in planar imaging.Keywords
This publication has 15 references indexed in Scilit:
- A New Architecture for Pixellated Solid State Gamma Camera Used in Nuclear MedicineIEEE Transactions on Nuclear Science, 2008
- Simulation and experimental results on monolithic CdZnTe gamma-ray detectorsIEEE Transactions on Nuclear Science, 2005
- GATE: a simulation toolkit for PET and SPECTPhysics in Medicine & Biology, 2004
- NUCAM3-a gamma camera based on segmented monolithic CdZnTe detectorsIEEE Transactions on Nuclear Science, 2002
- Medical applications of CdTe and CdZnTe detectorsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2001
- Applications of semiconductor detectors to nuclear medicineNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1999
- Signal-to-noise optimization of medical imaging systemsJournal of the Optical Society of America A, 1999
- Current state-of-the-art industrial and research applications using room-temperature CdTe and CdZnTe solid state detectorsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1996
- 1-D position sensitive single carrier semiconductor detectorsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1996
- Accelerated image reconstruction using ordered subsets of projection dataIEEE Transactions on Medical Imaging, 1994