The behaviour of a contactless conductivity detector was studied in its application to capillary electrophoresis employing quartz capillaries with inner diameters (i.d.) of 10 to 75 µm. The detector output signal was measured using KCl as the test electrolyte, within a KCl concentration range from 0 to 100 mmol/L, corresponding to specific conductivities from 0 to 1295 mS/m. When using capillaries with high inner diameters, then the signal‐to‐noise ratio is high in electrolytes of low conductivity. On the other hand, the use of capillaries with low i.d. values is useful in separations employing solutions of high conductivities. The advantages of capillaries with small i.d. values, combined with contactless conductivity detection, are demonstrated on separations of mixtures of inorganic ions and on separations of neutral mono‐ and disaccharides.Keywords
This publication has 14 references indexed in Scilit:
- Rapid monitoring of arrays of amino acids in clinical samples using capillary electrophoresis with contactless conductivity detectionJournal of Separation Science, 2010
- Ten years of axial capacitively coupled contactless conductivity detection for CZE – a reviewElectrophoresis, 2009
- Determination of ammonia, creatinine and inorganic cations in urine using CE with contactless conductivity detectionJournal of Separation Science, 2008
- A review of the recent achievements in capacitively coupled contactless conductivity detectionAnalytica Chimica Acta, 2008
- Contactless conductivity detection for microfluidics: Designs and applicationsTalanta, 2007
- Recent applications of conductivity detection in capillary and chip electrophoresisJournal of Separation Science, 2006
- Capacitively coupled contactless conductivity detection in capillary electrophoresisElectrophoresis, 2003
- Optimization of the high-frequency contactless conductivity detector for capillary electrophoresisElectrophoresis, 2002
- Contactless Conductivity Detector for Microchip Capillary ElectrophoresisAnalytical Chemistry, 2002
- Zone electrophoresis in open-tubular glass capillariesAnalytical Chemistry, 1981