Repeatability of non-invasive measurement of intracerebral pulse wave velocity using transcranial Doppler
Authors:
- 22 April 2005
- journal article
- Published by Portland Press Ltd. in Clinical Science
- Vol. 108 (5) , 433-439
Abstract
In the present study, the repeatability of three techniques for measuring peripheral PWV (pulse wave velocity) has been studied. A transcranial Doppler provided a wave reading from the middle cerebral artery. Using the transit time between the R-wave of an ECG and the ‘foot’ of this wave we were able to calculate a PWV (PWV-brain). An ear clip transducer provided a pressure wave reading (PWV-ear). A third pressure reading came from a Finapres transducer on the left middle finger (PWV-finger). The PWV was calculated as distance between two points/transit time of the pulse wave. Eleven volunteers had three sets of readings averaged for each technique taken in two separate sessions. There was good agreement between observers for the mean PWV values, and good agreement for mean results in different sessions. The RC%s (repeatability coefficient percentages) for between-observer repeatability in each session were good and approximately equivalent for PWV-finger (5–7%) and PWV-brain (5–7%). The repeatability of the PWV-ear measurement was less satisfactory (8–18%). The RC% for the same observer between sessions was less good, being 11% for the PWV-finger, 16–17% for PWV-brain and 11–19% for PWV-ear. The RC%s for the inter-session inter-observer measurements were between 10.7–12.1% for the PWV-finger, 14.7–19.5% for PWV-brain and 8.3–15% for PWV-ear. The transit time RC%s were lower in most measurements. The between-observer repeatability of all measures was satisfactory. Owing to the less good repeatability on different occasions, the use of PWV-brain and PWV-ear will depend on the magnitude of differences to be expected.Keywords
This publication has 10 references indexed in Scilit:
- Age-related changes in peripheral pulse timing characteristics at the ears, fingers and toesJournal of Human Hypertension, 2002
- Improvement in Blood Pressure, Arterial Stiffness and Wave Reflections With a Very-Low-Dose Perindopril/Indapamide Combination in Hypertensive PatientHypertension, 2001
- Transcranial Doppler ultrasoundBritish Medical Bulletin, 2000
- The effect of age on vascular compliance in man: which are the appropriate measures?Journal of Human Hypertension, 1999
- Systemic Arterial Compliance Early and Late after a First Acute Myocardial InfarctionCardiology, 1996
- Assessment of Arterial Distensibility by Automatic Pulse Wave Velocity MeasurementHypertension, 1995
- Atherosclerosis in the ascending aorta and risk of ischaemic strokeThe Lancet, 1995
- Arterial Wall Compliance in DiabetesDiabetic Medicine, 1992
- Stiffness of systemic arteries in patients with myocardial infarction. A noninvasive method to predict severity of coronary atherosclerosis.Circulation, 1989
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986