Lateral response heterogeneity of Bragg peak ionization chambers for narrow-beam photon and proton dosimetry
- 21 November 2017
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 62 (24) , 9189-9206
Abstract
Large area ionization chambers (LAICs) can be used to measure output factors of narrow beams. Dose area product measurements are proposed as an alternative to central-axis point dose measurements. Using such detectors requires detailed information on the uniformity of the response along the sensitive area. Eight LAICs were investigated in this study: four of type PTW-34070 (LAIC Thick ) and four of type PTW-34080 (LAIC Thin ). Measurements were performed in an x-ray unit using peak voltages of 100–200 kVp and a collimated beam of 3.1 mm (FWHM). The LAICs were moved with a step size of 5 mm to measure the chamber response at lateral positions. To account for beam positions where only a fraction of the beam impinged within the sensitive area of the LAICs, a corrected response was calculated which was the basis to calculate the relative response. The impact of a heterogeneous LAIC response, based on the obtained response maps was henceforth investigated for proton pencil beams and small field photon beams. A pronounced heterogeneity of the responses was observed in the investigated LAICs. The response of LAIC Thick generally decreased with increasing radius, resulting in a response correction of up to 5%. This correction was more pronounced and more diverse (up to 10%) for LAIC Thin . Considering a proton pencil beam the systematic offset for reference dosimetry was 2.4–4.1% for LAIC Thick and −9.5 to 9.4% for LAIC Thin . For relative dosimetry (e.g. integral depth-dose curves) systematic response variation by 0.8–1.9% were found. For a decreasing photon field size the systematic offset for absolute dose measurements showed a 2.5–4.5% overestimation of the response for 6 × 6 mm 2 field sizes for LAIC Thick . For LAIC Thin the response varied even over a range of 20%. This study highlights the need for chamber-dependent response maps when using LAICs for absolute and relative dosimetry with proton pencil beams or small photon beams.Keywords
Funding Information
- Christian Doppler Forschungsgesellschaft (CD10277401)
This publication has 32 references indexed in Scilit:
- Dosimetry intercomparison of four proton therapy institutions in Germany employing spot scanningZeitschrift für Medizinische Physik, 2017
- An overview on small-field dosimetry in photon beam radiotherapy: Developments and challengesJournal of Cancer Research and Therapeutics, 2017
- Recent developments in Geant4Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2016
- Detector to detector corrections: A comprehensive experimental study of detector specific correction factors for beam output measurements for small radiotherapy beamsMedical Physics, 2014
- Golden beam data for proton pencil-beam scanningPhysics in Medicine & Biology, 2012
- A procedure to determine the planar integral spot dose values of proton pencil beam spotsMedical Physics, 2012
- Determination of output factors for stereotactic radiosurgery beamsMedical Physics, 2009
- A new formalism for reference dosimetry of small and nonstandard fieldsMedical Physics, 2008
- The dose-area product, a new parameter for the dosimetry of narrow photon beamsZeitschrift für Medizinische Physik, 2006
- Geant4—a simulation toolkitNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2003