Pilot study of a new freely available computer-aided polyp detection system in clinical practice
Authors:
Open Access
- 11 May 2022
- journal article
- research article
- Published by Springer Nature in International Journal of Colorectal Disease
- Vol. 37 (6) , 1349-1354
Abstract
Purpose: Computer-aided polyp detection (CADe) systems for colonoscopy are already presented to increase adenoma detection rate (ADR) in randomized clinical trials. Those commercially available closed systems often do not allow for data collection and algorithm optimization, for example regarding the usage of different endoscopy processors. Here, we present the first clinical experiences of a, for research purposes publicly available, CADe system. Methods: We developed an end-to-end data acquisition and polyp detection system named EndoMind. Examiners of four centers utilizing four different endoscopy processors used EndoMind during their clinical routine. Detected polyps, ADR, time to first detection of a polyp (TFD), and system usability were evaluated (NCT05006092). Results: During 41 colonoscopies, EndoMind detected 29 of 29 adenomas in 66 of 66 polyps resulting in an ADR of 41.5%. Median TFD was 130 ms (95%-CI, 80–200 ms) while maintaining a median false positive rate of 2.2% (95%-CI, 1.7–2.8%). The four participating centers rated the system using the System Usability Scale with a median of 96.3 (95%-CI, 70–100). Conclusion: EndoMind’s ability to acquire data, detect polyps in real-time, and high usability score indicate substantial practical value for research and clinical practice. Still, clinical benefit, measured by ADR, has to be determined in a prospective randomized controlled trial.Keywords
Funding Information
- State government of Baden-Württemberg, Funding cluster “Forum Gesundheitsstandort Baden-Württemberg” (5409.0-001.01/15, 5409.0-001.01/15)
- Interdisziplinäres Zentrum für Klinische Forschung, Universitätsklinikum Würzburg (F-406, F-406)
- Bavarian Center for Cancer Research
- Universitätsklinikum Würzburg
This publication has 14 references indexed in Scilit:
- Performance of artificial intelligence in colonoscopy for adenoma and polyp detection: a systematic review and meta-analysisGastrointestinal Endoscopy, 2021
- Efficacy of Real-Time Computer-Aided Detection of Colorectal Neoplasia in a Randomized TrialGastroenterology, 2020
- Effect of a deep-learning computer-aided detection system on adenoma detection during colonoscopy (CADe-DB trial): a double-blind randomised studyThe Lancet Gastroenterology & Hepatology, 2020
- Detection of colorectal adenomas with a real-time computer-aided system (ENDOANGEL): a randomised controlled studyThe Lancet Gastroenterology & Hepatology, 2020
- Impact of a real-time automatic quality control system on colorectal polyp and adenoma detection: a prospective randomized controlled study (with videos)Gastrointestinal Endoscopy, 2020
- Study on detection rate of polyps and adenomas in artificial-intelligence-aided colonoscopySaudi Journal of Gastroenterology, 2020
- New artificial intelligence system: first validation study versus experienced endoscopists for colorectal polyp detectionGut, 2019
- Real-time automatic detection system increases colonoscopic polyp and adenoma detection rates: a prospective randomised controlled studyGut, 2019
- Deep Learning Localizes and Identifies Polyps in Real Time With 96% Accuracy in Screening ColonoscopyGastroenterology, 2018
- Performance measures for lower gastrointestinal endoscopy: a European Society of Gastrointestinal Endoscopy (ESGE) Quality Improvement InitiativeEndoscopy, 2017