Determination of Calcium and Magnesium by Chelometric Indicator Titrations with the Cupric Ion Electrode

Authors:

Abstract
A chelometric indicator titration technique using the cupric electrode was developed for application in the determination of Ca++ and Mg++ in estuarine waters and biological samples. This method has uses in both laboratory and field studies. In routine work. the lower limits of determination are about 0.2ppm for Ca++ and for Mg++. If the Mg++ ions are in solution with Ca++, the calcium ion concentration must be of the same order of magnitude or smaller than the Mg++ concentration. A comparison has been provided between the data obtained by the titration technique and those of atomic absorption measurements.