Contact angle and indentation velocity dependency for a resonance sensor—Evaluation on soft tissue silicone models
- 3 April 2013
- journal article
- Published by Taylor & Francis in Journal of Medical Engineering & Technology
- Vol. 37 (3) , 185-196
Abstract
Human tissue stiffness can vary due to different tissue conditions such as cancer tumours. Earlier studies show that stiffness may be detected with a resonance sensor that measures frequency shift and contact force at application. Through the frequency shift and the contact force, a tissue stiffness parameter can be derived. This study evaluated how the probe application angle and indentation velocity affected the results and determined the maximum parameter errors. The evaluation was made on flat silicone discs with specified hardness. The frequency shift, the force and the stiffness parameter all varied with contact angle and indentation velocity. A contact angle of ≤10° was acceptable for reliable measurements. A low indentation velocity was recommended. The maximum errors for the system were <1.1% of the measured values. It was concluded that contact angle and indentation velocity have to be considered in the clinical setting. The angular dependency is especially important in clinical use for studying stiffness of human soft tissue, e.g. in prostate cancer diagnosis.Keywords
This publication has 13 references indexed in Scilit:
- Tactile resonance sensors in medicineJournal of Medical Engineering & Technology, 2009
- Explanatory models for a tactile resonance sensor system—elastic and density-related variations of prostate tissuein vitroPhysiological Measurement, 2008
- Resonance sensor measurements of stiffness variations in prostate tissuein vitro—a weighted tissue proportion modelPhysiological Measurement, 2006
- Prostate cancer detection with an improved resonance sensor system: parameter evaluation in a silicone model and on human prostate tissue in vitroMedical & Biological Engineering & Computing, 2006
- Prostate tissue stiffness as measured with a resonance sensor system: a study on silicone and human prostate tissue in vitroMedical & Biological Engineering & Computing, 2006
- Objective evaluation of liver consistency to estimate hepatic fibrosis and functional reserve for hepatectomy1Journal of the American College of Surgeons, 2000
- Palpation of the skin with a robot finger: an attempt to measure skin stiffness with a probe loaded with a newly developed tactile vibration sensor and displacement sensorSkin Research and Technology, 1999
- The stiffness of lymph nodes containing lung carcinoma metastasesCancer, 1997
- New tactile sensor like the human hand and its applicationsSensors and Actuators A: Physical, 1992
- Impression technique for the assessment of oedemaMedical & Biological Engineering & Computing, 1991