A New High Resolution Stopped-Flow Apparatus
Author:
- Jack I. MorrowJMJack I. Morrow
- 1 January 1970
- journal article
- research article
- Published by Taylor & Francis in Instrumentation Science & Technology
- Vol. 2 (4) , 375-387
Abstract
A new stopped-flow apparatus for the study of fast reactions is described. It employs a porting arrangement (through which the reservoir and drive syringes communicate) which minimizes bubble problems and permits solutions to be changed in less than 1 min. A dual path length flow cell is also described (10 mm light path along the line of flow and a 2 mm light path normal to the line of flow). Under a driving pressure of 60 lb/in2† the measured dead times were about 4.0 and 2.0 msec for the long and the short light path modes, respectively. The capacity of each of the drive syringes is 2 ml, and less than 0.1 ml of each reagent need be expended per experiment. When the 10 mm light path mode is employed, the volume under observation is 0.04 ml.Keywords