Smart laser osteotomy: integrating a pulsed 1064nm fiber laser into the sample arm of a fiber optic 1310nm OCT system for ablation monitoring
- 19 November 2018
- journal article
- research article
- Published by Optica Publishing Group in Biomedical Optics Express
- Vol. 9 (12) , 6374-6387
Abstract
Real-time depth metrology during material removal via laser ablation is useful in many forms of laser machining. Until now, coaxial optical coherence tomography (OCT) metrology was achieved by the coupling of an OCT imaging beam and ablating beams using a dichroic filter. We present an alternative design with all fiber delivery which is more suitable for surgical laser ablation applications. The novel system design integrates a high peak-power pulsed Yb-doped fiber laser (1064nm) coupled directly into the sample arm of a swept-source OCT system (λc = 1310nm). We measured the OCT signal degradation due to dispersion and attenuation through the ablation fiber laser cavity. Ablation progression is measured in real-time using M-mode OCT. Mean depth targeting error was found to range from 10µm to 80µm in phantom ablation experiments and 21µm to 60µm in bone ablation. A number of issues have been solved, including point-spread function (PSF) peak broadening due to signal delay and dispersion, high bending loss due to dissimilar fiber used throughout the design, and problems due to the extremely high ablation power to swept-source power ratio (> 2×104 peak to average power). To our knowledge, this is the first demonstration of thermal-mediated laser ablation drilling integrated with coaxial OCT imaging through a single-mode, single-cladded output fiber, without using dichroic beam splitters or free-space optic filters anywhere in the optical path and with this high ablation laser power to OCT source power ratio. The removal of bulk optics compared to existing designs opens a new path for compact integration of the entire system. Also, since the ablation laser and OCT feedback system exist along the same fiber path, the need for maintenance and repair are greatly reduced since spatial beam alignment and potential open-air contamination of optical surfaces are virtually eliminated. We believe that this integrated system is a great candidate for adoption in depth-controlled surgical ablation applications.Keywords
Funding Information
- Canada Foundation for Innovation (CFI)
- Natural Sciences and Engineering Research Council of Canada (NSERC)
This publication has 15 references indexed in Scilit:
- Laser tissue coagulation and concurrent optical coherence tomography through a double-clad fiber couplerBiomedical Optics Express, 2015
- Real‐time guidance of thermal and ultrashort pulsed laser ablation in hard tissue using inline coherent imagingLasers in Surgery and Medicine, 2012
- Planning and simulation of microsurgical laser bone ablationInternational Journal of Computer Assisted Radiology and Surgery, 2009
- Ultrahigh-speed optical frequency domain imaging and application to laser ablation monitoringApplied Physics Letters, 2006
- Noninvasive Measurement of Ablation Crater Size and Thermal Injury after CO2 Laser in the Vocal Cord with Optical Coherence TomographyOtolaryngology–Head and Neck Surgery, 2006
- In-situ Observation of Tissue Laser Ablation Using Optical Coherence TomographyOptical and Quantum Electronics, 2005
- Ultrahigh-resolution, high-speed, Fourier domain optical coherence tomography and methods for dispersion compensationOptics Express, 2004
- Real-Time Optical Coherence Tomography for Minimally Invasive Imaging of Prostate AblationComputer Aided Surgery, 2001
- High-Resolution Optical Coherence Tomography-Guided Laser Ablation of Surgical TissueJournal of Surgical Research, 1999
- Photophysical processes in recent medical laser developments: A reviewLasers in Medical Science, 1986