A 9-independent-leads ECG system from 10 electrodes: A practice preserving WCT-less true unipolar ECG system
- 1 October 2015
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- Vol. 110, 1-4
Abstract
Since the invention of the 12-lead Electrocardiography (ECG) it is taught that only eight out of the twelve signals recordable from the nine exploring electrodes are linearly independent. It is also accepted practice the use of other four linearly dependent signals to add diagnostic efficacy. Unfortunately, quality and usefulness of the 12-lead signals is directly correlated with electrode placement precision and concerns have been raised around intentional misplacements of limb electrodes also for resting ECG. In this paper we show that with a simple modification to the recording front-end, nine independent signals (one per recording electrode) can be recorded. These signals can be used to reconstruct precisely the 12-lead signals (without information loss) and a plethora of other linearly dependent new signals that can open an unexplored frontier in cardiac illnesses diagnosis. While we need to carefully evaluate the clinical implications of our findings, we present here the proposed circuit and a case study (healthy subject) showing the number of signals achievable and we quantify the effect upon the ECG signals of one of the most common intentional misplacement of electrodes.Keywords
This publication has 12 references indexed in Scilit:
- Towards true unipolar ECG recording without the Wilson central terminal (preliminary results)Physiological Measurement, 2013
- Unipolar ECG circuits: Towards more precise cardiac event identification2013
- Does modifying electrode placement of the 12 lead ECG matter in healthy subjects?International Journal of Cardiology, 2011
- An ultra-high input impedance ECG amplifier for long-term monitoring of athletesMedical Devices: Evidence and Research, 2010
- A mobile EEG system with dry electrodes2008
- Practice Standards for Electrocardiographic Monitoring in Hospital SettingsCirculation, 2004
- Reduction of Interference Due to Common Mode Voltage in Biopotential AmplifiersIEEE Transactions on Biomedical Engineering, 1983
- MEASUREMENT OF THE ERROR IN WILSON'S CENTRAL TERMINAL: AN ACCURATE DEFINITION OF UNIPOLAR LEADSHeart, 1959
- The zero of potential: A persistent errorAmerican Heart Journal, 1955
- The Problem of Adjusting the Wilson Central Terminal to a Zero of Potential in the Living Human SubjectCirculation Research, 1955