Head-and-neck IMRT treatments assessed with a Monte Carlo dose calculation engine
Authors:
- J SecoJSJ Seco
- E AdamsEAE Adams
- M BidmeadMBM Bidmead
- M PartridgeMPM Partridge
- F VerhaegenFVF Verhaegen
- 18 February 2005
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 50 (5) , 817-830
Abstract
IMRT is frequently used in the head-and-neck region, which contains materials of widely differing densities (soft tissue, bone, air-cavities). Conventional methods of dose computation for these complex, inhomogeneous IMRT cases involve significant approximations. In the present work, a methodology for the development, commissioning and implementation of a Monte Carlo (MC) dose calculation engine for intensity modulated radiotherapy (MC-IMRT) is proposed which can be used by radiotherapy centres interested in developing MC-IMRT capabilities for research or clinical evaluations. The method proposes three levels for developing, commissioning and maintaining a MC-IMRT dose calculation engine: (a) development of a MC model of the linear accelerator, (b) validation of MC model for IMRT and (c) periodic quality assurance (QA) of the MC-IMRT system. The first step, level (a), in developing an MC-IMRT system is to build a model of the linac that correctly predicts standard open field measurements for percentage depth-dose and off-axis ratios. Validation of MC-IMRT, level (b), can be performed in a rando phantom and in a homogeneous water equivalent phantom. Ultimately, periodic quality assurance of the MC-IMRT system is needed to verify the MC-IMRT dose calculation system, level (c). Once the MC-IMRT dose calculation system is commissioned it can be applied to more complex clinical IMRT treatments. The MC-IMRT system implemented at the Royal Marsden Hospital was used for IMRT calculations for a patient undergoing treatment for primary disease with nodal involvement in the head-and-neck region (primary treated to 65 Gy and nodes to 54 Gy), while sparing the spinal cord, brain stem and parotid glands. Preliminary MC results predict a decrease of approximately 1-2 Gy in the median dose of both the primary tumour and nodal volumes (compared with both pencil beam and collapsed cone). This is possibly due to the large air-cavity (the larynx of the patient) situated in the centre of the primary PTV and the approximations present in the dose calculation.Keywords
This publication has 19 references indexed in Scilit:
- Clinical implementation of dynamic and step-and-shoot IMRT to treat prostate cancer with high risk of pelvic lymph node involvementRadiotherapy and Oncology, 2004
- A Monte Carlo based phase space model for quality assurance of intensity modulated radiotherapy incorporating leaf specific characteristicsMedical Physics, 2002
- Energy spectra, angular spread, fluence profiles and dose distributions of 6 and 18 MV photon beams: results of Monte Carlo simulations for a Varian 2100EX acceleratorPhysics in Medicine & Biology, 2002
- A method of simulating dynamic multileaf collimators using Monte Carlo techniques for intensity-modulated radiation therapyPhysics in Medicine & Biology, 2001
- Accurate condensed history Monte Carlo simulation of electron transport. I. EGSnrc, the new EGS4 versionMedical Physics, 2000
- Dose calculations for external photon beams in radiotherapyPhysics in Medicine & Biology, 1999
- The flatness of Siemens linear accelerator x‐ray fieldsMedical Physics, 1999
- A technique for the quantitative evaluation of dose distributionsMedical Physics, 1998
- 3D electron dose calculation using a Voxel based Monte Carlo algorithm (VMC)Medical Physics, 1996
- A Monte Carlo study of accelerator head scatterMedical Physics, 1994