Elastohydrodynamic lubrication analysis of hip implants with ultra high molecular weight polyethylene cups under transient conditions
Authors:
- 1 July 2003
- journal article
- research article
- Published by SAGE Publications in Proceedings of the Institution of Mechanical Engineers, Part C: Journal of Mechanical Engineering Science
- Vol. 217 (7) , 767-777
Abstract
The transient variation of both the load and speed experienced during walking in an elastohydrodynamic lubrication (EHL) analysis for artificial hip joints employing an ultra high molecular weight polyethylene (UHMWPE) acetabular cup against either a metallic or ceramic femoral head was considered in this study. A general numerical procedure to solve the transient EHL problem in spherical ball-in-socket coordinates, developed in a previous study by Jalali-Vahid and Jin in 2002, was applied under three specific conditions experienced during typical gait cycles, including speed reversal, a sudden load increase and a sudden load decrease. The predicted minimum film thickness was found to stay remarkably constant and similar to that prior to the change in either the load or the angular velocity, despite a large change in these operating conditions. This was attributed to the remarkably effective squeeze-film action of preserving and maintaining the lubricating film developed before the transient variations in either the load or speed. It is therefore possible to neglect the effect of these specific transient variations of load and speed under physiological walking conditions considered in the present study on the predicted film thickness in hip implants with UHMWPE cups.Keywords
This publication has 16 references indexed in Scilit:
- “In vivo” determination of hip joint separation and the forces generated due to impact loading conditionsJournal of Biomechanics, 2001
- Elastohydrodynamic lubrication analysis of UHMWPE hip joint replacements2000
- Wear and Lubrication of Metal-on-Metal Hip ImplantsClinical Orthopaedics and Related Research, 1999
- Numerical Analysis of Time-Varying Fluid Film Thickness in Metal-Metal Hip Implants in Simulator Tests1998
- Fluid film lubrication in natural hip jointsTribology Series, 1993
- The Minimum Film Thickness in Line Contacts During Reversal of EntrainmentJournal of Tribology, 1993
- Numerical analysis of elastohydrodynamic squeeze film lubrication of total hip prosthesesWear, 1990
- Calculation of Elastohydrodynamic Lubrication Film Thickness for Hip Prostheses During Normal WalkingTribology Transactions, 1990
- Reactions of the articular capsule to wear products of artificial joint prosthesesJournal of Biomedical Materials Research, 1977
- The Frictional Behavior of Human Synovial Joints—Part II: Artificial JointsJournal of Lubrication Technology, 1975