A study of the long-term precision of dual-energy X-ray absorptiometry bone densitometers and implications for the validity of the least-significant-change calculation
- 30 November 2006
- journal article
- research article
- Published by Springer Nature in Osteoporosis International
- Vol. 18 (4) , 513-523
Abstract
Short-term precision is often quoted and used as the most important performance parameter of a dual-energy X-ray absorptiometry (DXA) scanner; however, long-term precision has a more profound impact on patient monitoring. Long-term precision refers to the combination of in-vivo precision errors and long-term equipment stability. To monitor long-term equipment stability, a phantom was designed with four inserts ranging in bone-mineral density from 0.5 to 3.3 g/cm2. This phantom was used to monitor the equipment stability of four modern fan-beam densitometers, two each from Hologic and GE/Lunar, over a 4-year period. Manufacturer-recommended quality assurance (QA) procedures were performed, and the scanners stayed within manufacturer-specified tolerances throughout the study. During the 4-year period, the Hologic scanners were observed to cause clinically insignificant BMD shifts (maximum of 0.34%), whereas the GE/Lunar scanners revealed BMD shifts that were clinically significant (1.5% and 2.1%). As a result, using least-significant-change (LSC) calculations based only on short-term in-vivo precision studies for monitoring patients is not valid for the two GE/Lunar densitometers due to the poorer long-term stability they exhibited.Keywords
This publication has 13 references indexed in Scilit:
- Official Positions of the International Society for Clinical Densitometry and Executive Summary of the 2005 Position Development ConferenceJournal of Clinical Densitometry, 2006
- Reproducibility of bone mineral density measurements using dual X-ray absorptiometry in daily clinical practiceOsteoporosis International, 2005
- Need for Precision Studies at Individual Institutions and Assessment of Size of Regions of Interest on Serial DXA ScansJournal of Clinical Densitometry, 2003
- What Is the Role of Serial Bone Mineral Density Measurements in Patient Management?Journal of Clinical Densitometry, 2002
- Importance of Precision in Bone Density MeasurementsJournal of Clinical Densitometry, 2001
- Instrument performance in bone density testing at five Australian centersAustralian and New Zealand Journal of Medicine, 1997
- A Preliminary Evaluation of the Lunar Expert-XL for Bone Densitometry and Vertebral MorphometryJournal of Bone and Mineral Research, 1997
- Long-term performance in vitro and in vivo of dual-energy X-ray absorptiometryClinical Rheumatology, 1995
- Precision of dual x-ray absorptiometry and peripheral computed tomography using mobile densitometry unitsCalcified Tissue International, 1994
- Longitudinal precision of dual-energy X-ray absorptiometry in a multicenter studyJournal of Bone and Mineral Research, 1991