High sensitivity gas pressure sensor based on different inner diameter quartz capillary cascading and Vernier effect
- 1 August 2022
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 93 (8) , 085002
Abstract
In this paper, a highly sensitive optical fiber gas pressure sensor is proposed and experimentally verified. The sensor is composed of two Fabry–Pérot (F-P) cavities, and two F-P cavities are fabricated by a single-mode fiber and two quartz capillaries with different inner diameters splicing. Among them, the small inner diameter capillary is used as a gas channel connecting the large inner diameter capillary and the external environment. The manufacturing process of the sensor only involves capillary cleaver and splicing and does not involve other complex manufacturing technologies. By correctly adjusting the length of the two quartz capillaries, when the free spectral range of the two F–P cavities is very close, the optical Vernier effect will be observed and used as a sensitive probe for detecting gas pressure. The experimental results show that, in the pressure range of 0–0.8 MPa, the gas pressure sensitivity of the sensor reaches −81.73 nm/MPa with a linearity of 99.7%, and the temperature cross-sensitivity is only 1.82 kPa/°C. Due to its easy manufacture, high sensitivity, compact structure, and small volume, the sensor has become one of the preferred structures for large-scale use in the field of gas sensing.Keywords
Funding Information
- The Guiding Project of Scientific Research Plan of the Department of Education of Hubei Province Hubei Provincial Department of Education (B2020136)
- The Middle-Aged and Young Science and Technology Innovation Team of Hubei Province Universities Hubei Provincial Department of Education (T2020014)
- the Fund of Graduate Innovation Research of Hubei Normal University Hubei Normal University (20210111)
This publication has 27 references indexed in Scilit:
- Simultaneous measurement of transverse load and temperature using hybrid structured fiber-optic Fabry–Perot interferometerScientific Reports, 2017
- Long Period Fiber Grating Inscribed in Hollow-Core Photonic Bandgap Fiber for Gas Pressure SensingIEEE Photonics Journal, 2017
- High-Sensitivity Gas-Pressure Sensor Based on Fiber-Tip PVC Diaphragm Fabry–Pérot InterferometerJournal of Lightwave Technology, 2017
- Silk fibroin diaphragm-based fiber-tip Fabry-Perot pressure sensorOptics Express, 2016
- Fiber Fabry–Pérot Interferometer for Measurement of Gas Pressure and TemperatureJournal of Lightwave Technology, 2016
- High-Sensitivity Gas Pressure Sensor Based on Fabry–Pérot Interferometer With a Side-Opened Channel in Hollow-Core Photonic Bandgap FiberIEEE Photonics Journal, 2015
- Fiber-tip gas pressure sensor based on dual capillariesOptics Express, 2015