Analysis, design and control of a transcutaneous power regulator for artificial heart
- 1 June 2008
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE) in 2008 IEEE Power Electronics Specialists Conference
- No. 02759306,p. 1833-1838
Abstract
Based on commonly used parameters for a generic-transcutaneous transformer model, a remote power supply using resonant topology for artificial heart is analyzed and designed for easy controllability and high efficiency. Primary and secondary windings of the transcutaneous transformer are positioned outside and inside human body respectively for energy transfer. The two large leakage inductances and the mutual inductance of the transformer are varying parameters of the coupling-coefficient which varies with transformer alignment and gap due to external positioning. Varying resonant-frequency resonant-tank circuits are formed using the transformer inductors and external capacitors to obtain a load insensitive frequency for the voltage transfer function at given range of coupling coefficients and loads. Previous researches usually use frequency modulation which may require a wide control frequency range well above the load insensitive frequency. In this paper, fundamental frequency study of the input-to-output voltage transfer function is carried out. Using the proposed control method, the switching frequency can be locked at just above the load insensitive frequency at heavy load for best efficiency. Specifically, above resonant operation in driving the resonant circuits when varying the coupling-coefficient is maintained using a digital-phase-lock-loop (PLL) technique to achieve zero-voltage switching of a full-bridge switches configuration which is also programmed to provide pulse-width-modulation (PWM) in controlling the output voltage. A prototype transcutaneous power regulator is built and found to have good efficiency and regulation in responding to changing alignment or gap of the transcutaneous transformer, load and input voltage dynamically.Keywords
This publication has 22 references indexed in Scilit:
- Feedback Analysis and Design of RF Power Links for Low-Power Bionic SystemsIEEE Transactions on Biomedical Circuits and Systems, 2007
- Conception of an emitter for transcutancous power and high-speed data transmission2005
- The design of core-type transcutaneous energy transmission systems for artificial heart2005
- Efficiency improvement and in vivo estimation of externally-coupled transcutaneous energy transmission system for a totally implantable artificial heart2002
- An improved transmission energy transformer for a non invasive rechargeable battery to artificial organs2002
- Artificial hearts, batteries, and electric vehiclesIEEE Aerospace and Electronic Systems Magazine, 2002
- A new structure transcutaneous transformer for artificial heart systemIEEE Transactions on Magnetics, 1999
- An energy transmission system for an artificial heart using leakage inductance compensation of transcutaneous transformerIEEE Transactions on Power Electronics, 1998
- Transcutaneous DC-DC converter for totally implantable artificial heart using synchronous rectifierIEEE Transactions on Magnetics, 1996
- Investigation of coil geometry for transcutaneous energy transmission for artificial heartIEEE Transactions on Magnetics, 1992