Alumina-on-alumina ceramic versus metal-on-highly cross-linked polyethylene bearings in total hip arthroplasty: a comparative study
Open Access
- 1 November 2009
- journal article
- research article
- Published by Springer Nature in International Orthopaedics
- Vol. 34 (8) , 1129-1135
Abstract
The aim of the study was to evaluate the reliability and durability of alumina-on-alumina ceramic in comparison to metal-on-highly cross-linked polyethylene (CoCr/HXLPE) bearing couples. This prospective randomised study involved 150 patients (157 hips). All patients (mean age: 54.7 years) obtained an identical fibre metal midcoat femoral stem and fibre metal-coated acetabular shell. In 78 patients (82 hips) we used alumina, while in 72 patients (75 hips) metal-polyethylene bearing couples were used. During a mean 50.4-month follow-up period (51 ± 8 alumina and 50 ± 8.9 metal-polyethylene) no statistically significant changes in clinical and radiographic parameters were noted between the two groups. There was no ceramic breakage and no need for revision surgery due to the ceramic liner. The alumina bearing couples proved to be as reliable as CoCr/HXLPE.Keywords
This publication has 26 references indexed in Scilit:
- A systematic review of radiological outcomes of highly cross-linked polyethylene versus conventional polyethylene in total hip arthroplastyInternational Orthopaedics, 2009
- Ceramic-on-Ceramic Total Hip Arthroplasty: UpdateThe Journal of Arthroplasty, 2008
- “Severe” wear challenge to 36 mm mechanically enhanced highly crosslinked polyethylene hip linersJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2008
- The Role of Stripe Wear in Causing Acoustic Emissions From Alumina Ceramic-On-Ceramic BearingsThe Journal of Arthroplasty, 2007
- PRESIDENTIAL GUEST ADDRESS: The Swedish Hip RegistryClinical Orthopaedics and Related Research, 2005
- Analysis of the kinematics of different hip simulators used to study wear of candidate materials for the articulation of total hip arthroplastiesJournal of Orthopaedic Research, 1998
- Loci of movement of selected points on the femoral head during normal gaitThe Journal of Arthroplasty, 1996
- Fracture of the femoral head after ceramic-on-polyethylene total hip arthroplastyThe Journal of Arthroplasty, 1995
- Prognosis of total hip replacement in Sweden: Follow-up of 92,675 operations performed 1978–1990Acta Orthopaedica, 1993
- Role of Ceramic ImplantsClinical Orthopaedics and Related Research, 1992